锥形束CT在有症状全膝关节置换术后评估中的作用:20例报告
本研究对20例有症状的全膝关节置换术后患者行锥形束CT检查,由三名独立观察者测量髌骨倾斜角及股骨/胫骨假体旋转对位,并用四级评分评估金属伪影对假体周围骨与肌腱-肌肉结构识别的影响;结果显示测量可重复性很高,假体远端部分结构无法识别,而近端及胫骨柄近/中三分之一结构可良好识别。
要点
- 锥形束CT测量髌骨倾斜角及股骨/胫骨假体旋转对位具有很高的观察者内和观察者间一致性。
- 股骨假体远端部分的骨与肌腱-肌肉结构因金属伪影无法识别。
- 股骨假体近端及胫骨柄近/中三分之一的骨与肌腱-肌肉结构可良好识别。
核心结论
- 锥形束CT是有效工具,可重复测量髌骨倾斜角及股骨/胫骨假体旋转对位,并能在金属伪影干扰较小的情况下分析骨与肌腱-肌肉结构。
研究设计与样本
- 对20例有症状的全膝关节置换术后患者行锥形束CT检查,由三名独立观察者测量髌骨倾斜角及股骨/胫骨假体旋转对位,并以四级评分判断各结构的金属伪影程度。
关键影像表现/征象
- 股骨假体远端部分的骨与肌腱-肌肉结构无法识别。
- 股骨假体近端及胫骨柄近/中三分之一的骨与肌腱-肌肉结构可良好识别。
诊断标准与关键数值
- 髌骨倾斜角及假体旋转对位的观察者内一致性组内相关系数为0.895-0.975。
- 髌骨倾斜角及假体旋转对位的观察者间一致性组内相关系数为0.891-0.948。
- 采用四级评分,从「许多金属伪影」(结构无法识别)到「无金属伪影」(结构可完美识别),用于判断每一结构。
临床应用与阅片要点
- 锥形束CT可在旋转轴平面上进行简便且准确的测量,不受金属伪影所致图像质量下降的负担。
- 评估股骨假体远端部分时,骨与肌腱-肌肉结构因金属伪影无法识别。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The aims of this study were to evaluate the efficacy of cone beam CT (CBCT) in the study of the patellar tilt angle and rotational alignment of the femoral/tibial component after total knee arthroplasty and to estimate how metallic artefacts impaired detection of periprosthetic bone structures and bordering tendon-muscle structures.
METHODS: 20 symptomatic total knee arthroplasties were examined using CBCT by three independent observers. The patellar tilt angle and rotational alignment of femoral and tibial components were measured in relation to the femoral flange, transepicondylar axis and tibial tuberosity, respectively. A four-score scale, ranging from "many metallic artefacts" (the structure cannot be identified) to "no metallic artefacts" (the structure can be perfectly identified), was used to judge every structure.
RESULTS: The patellar tilt angle and rotational alignment of the prosthetic components showed very high intra- and interobserver agreements (intraclass correlation coefficient values 0.895-0.975 and 0.891-0.948, respectively). Bone and tendon-muscle structures cannot be identified in the distal part of the femoral component, whereas they can be well identified in the proximal part of the femoral component and in the proximal/middle third of the tibial stem.
CONCLUSION: CBCT was an effective tool, providing reproducible measurements of the patellar tilt angle and the rotational alignment of the femoral/tibial component. Furthermore, it allowed bone and tendon-muscle structures analysis with little impediments from metal artefacts. Advances in knowledge: CBCT allows easy and accurate measurements on the rotational axial plane, unburdened by image quality impairment due to metal artefacts.