模拟运动模糊对全视野数字乳腺X线摄影病灶检出性能的影响

The impact of simulated motion blur on lesion detection performance in full-field digital mammography.
PubMed 文献 · Br J Radiol · 2017-06-16

研究纳入7名阅片者(报告经验15±5年)评估248例(62例恶性肿块、62例恶性微钙化、124例正常),在无模糊(0 mm)及模拟模糊(0.7、1.5 mm)三种条件下行自由响应观察者研究,采用等权重刀切备择自由响应受试者操作特征分析。结果显示肿块与微钙化的检出性能均随模拟模糊幅度增加而显著下降,作者认为这些假阴性判断可能对临床实践有影响。

要点

  • 模拟运动模糊幅度增加时,肿块与微钙化的检出性能指标均下降。
  • 肿块检出在0.0与0.7 mm、0.0与1.5 mm间有统计学差异。
  • 微钙化检出在所有模糊幅度配对间均有统计学差异。

核心结论

  • 数学模拟的运动模糊导致病灶检出性能出现统计学显著下降,这些假阴性判断可能对临床实践有影响。

研究设计与样本

  • 自由响应观察者研究,7名阅片者(报告经验15±5年)评估248例(62例恶性肿块、62例恶性微钙化、124例正常),异常病例经活检证实,采用数学模拟提供像素位移以模拟运动模糊,比较无模糊(0 mm)与0.7、1.5 mm模拟模糊三种条件。

诊断标准与关键数值

  • 以等权重刀切备择自由响应受试者操作特征分析为品质因数,检验α设为0.05以控制I型错误概率;肿块检出性能总体差异F(2,22)=6.01、p=0.0084,微钙化检出性能总体差异F(2,49)=23.14、p<0.0001。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: Motion blur is a known phenomenon in full-field digital mammography, but the impact on lesion detection is unknown. This is the first study to investigate detection performance with varying magnitudes of simulated motion blur.

METHODS: 7 observers (15 ± 5 years' reporting experience) evaluated 248 cases (62 containing malignant masses, 62 containing malignant microcalcifications and 124 normal cases) for 3 conditions: no blurring (0 mm) and 2 magnitudes of simulated blurring (0.7 and 1.5 mm). Abnormal cases were biopsy proven. Mathematical simulation was used to provide a pixel shift in order to simulate motion blur. A free-response observer study was conducted to compare lesion detection performance for the three conditions. The equally weighted jackknife alternative free-response receiver operating characteristic was used as the figure of merit. Test alpha was set at 0.05 to control probability of Type I error.

RESULTS: The equally weighted jackknife alternative free-response receiver operating characteristic analysis found a statistically significant difference in lesion detection performance for both masses [F(2,22) = 6.01, p = 0.0084] and microcalcifications [F(2,49) = 23.14, p < 0.0001]. The figures of merit reduced as the magnitude of simulated blurring increased. Statistical differences were found between some of the pairs investigated for the detection of masses (0.0 vs 0.7 and 0.0 vs 1.5 mm) and all pairs for microcalcifications (0.0 vs 0.7, 0.0 vs 1.5 and 0.7 vs 1.5 mm). No difference was detected between 0.7 and 1.5 mm for masses.

CONCLUSION: The mathematical simulation of motion blur caused a statistically significant reduction in lesion detection performance. These false-negative decisions could have implications for clinical practice. Advances in knowledge: This research demonstrates for the first time that motion blur has a negative and statistically significant impact on lesion detection performance in digital mammography.

来源:Br J Radiol,查看原文。版权归原作者所有。