文献速递乳腺MR

压缩感知用于乳腺MRI:解决空间分辨率与时间分辨率之间的权衡

Compressed Sensing for Breast MRI: Resolving the Trade-Off Between Spatial and Temporal Resolution.
PubMed 文献 · Invest Radiol · 2017-01-01

研究纳入30例女性,评估原型压缩感知容积内插屏气检查(CS-VIBE)序列用于超快乳腺MRI,并与超快TWIST序列及常规VIBE-DIXON序列比较。CS-VIBE在相同采集时间内获得更高层间空间分辨率且平面内图像质量无显著差异(P=0.260),病灶形态学评估与扫描时间长18倍的VIBE-DIXON相当,敏感度(CS-VIBE 84.0% vs VIBE-DIXON 92.0%,P=0.500)与特异度(CS-VIBE 60.0% vs VIBE-DIXON 55.4%,P=0.327)无显著差异。

要点

  • CS-VIBE在相同采集时间内较TWIST获得更高层间空间分辨率,平面内图像质量无显著差异(P=0.260)。
  • CS-VIBE病灶形态学评估与扫描时间91秒的VIBE-DIXON相当,后者采集时间约为CS-VIBE的18倍。
  • CS-VIBE敏感度84.0% vs VIBE-DIXON 92.0%(P=0.500),特异度60.0% vs 55.4%(P=0.327),均无显著差异。

核心结论

  • 压缩感知容积内插屏气检查(CS-VIBE)可在与TWIST相当的平面内图像质量下提高超快动态增强乳腺MRI的层间空间分辨率,病灶形态学评估与扫描时间长18倍的VIBE-DIXON相当,从而支持超快乳腺MRI的3D评估。

研究设计与样本

  • 2016年4月至2016年10月纳入30例女性,以CS-VIBE序列替代混合乳腺MRI方案中的常规超快TWIST序列,所有扫描在3T MAGNETOM Skyra系统(Siemens Healthcare,Erlangen,Germany)上使用16通道双侧乳腺线圈完成;开展两项阅读者研究共涉及5名阅读者,第一项研究对23例有既往TWIST MRI的女性独立评分CS-VIBE与TWIST的轴位整体图像质量及伪影,第二项研究对16例有病灶女性的31个病灶(5个恶性、26个良性)按BI-RADS MRI词典独立描述CS-VIBE与VIBE-DIXON的病灶特征并给出病灶显著性评分。

关键影像表现/征象

  • CS-VIBE较TWIST显示略多的呼吸和搏动伪影,TWIST较CS-VIBE显示略多的运动伪影、折叠伪影和鬼影伪影。
  • CS-VIBE与VIBE-DIXON的阅者内一致性为轻度至几乎完美,且一般高于阅者间一致性。

诊断标准与关键数值

  • 以VIBE-DIXON为对照,CS-VIBE诊断敏感度为84.0%(VIBE-DIXON为92.0%,P=0.500),特异度为60.0%(VIBE-DIXON为55.4%,P=0.327)。
  • 病灶特征按BI-RADS MRI词典描述,并另给出病灶显著性评分。
  • 伪影存在情况采用5分制评分。

临床应用与阅片要点

  • CS-VIBE在保持超快时间分辨率(每期容积小于5秒)的同时提高层间空间分辨率,可用于超快乳腺MRI的3D评估。
  • 读片时需注意CS-VIBE较TWIST略多的呼吸和搏动伪影。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: Ultrafast dynamic contrast-enhanced magnetic resonance imaging of the breast enables assessment of the contrast inflow dynamics while providing images with diagnostic spatial resolution. However, the slice thickness of common ultrafast techniques still prevents multiplanar reconstruction. In addition, some temporal blurring of the enhancement characteristics occurs in case view-sharing is used. We evaluate a prototype compressed-sensing volume-interpolated breath-hold examination (CS-VIBE) sequence for ultrafast breast MRI that improves through plane spatial resolution and avoids temporal blurring while maintaining an ultrafast temporal resolution (less than 5 seconds per volume). Image quality (IQ) of the new sequence is compared with an ultrafast view-sharing sequence (time-resolved angiography with interleaved stochastic trajectories [TWIST]), and assessment of lesion morphology is compared with a regular T1-weighted 3D Dixon sequence (VIBE-DIXON) with an acquisition time of 91 seconds.

MATERIALS AND METHODS: From April 2016 to October 2016, 30 women were scanned with the CS-VIBE sequence, replacing the routine ultrafast TWIST sequence in a hybrid breast MRI protocol. The need for informed consent was waived. All MRI scans were performed on a 3T MAGNETOM Skyra system (Siemens Healthcare, Erlangen, Germany) using a 16-channel bilateral breast coil. Two reader studies were conducted involving 5 readers. In the first study, overall IQ of CS-VIBE and TWIST in the axial plane was independently rated for 23 women for whom prior MRI examinations with TWIST were available. In addition, the presence of several types of artifacts was rated on a 5-point scale. The second study was conducted in women (n = 16) with lesions. In total, characteristics of 31 lesions (5 malignant and 26 benign) were described independently for CS-VIBE and VIBE-DIXON, according to the BI-RADS MRI-lexicon. In addition, a lesion conspicuity score was given.

RESULTS: Using CS-VIBE, a much higher through-plane spatial resolution was achieved in the same acquisition time as with TWIST, without affecting in-plane IQ (P = 0.260). Time-resolved angiography with interleaved stochastic trajectories showed slightly more motion artifacts and infolding and ghosting artifacts compared with CS-VIBE, whereas CS-VIBE showed more breathing and pulsation artifacts. For morphologic assessment, intrareader agreement between CS-VIBE and the more time-consuming VIBE-DIXON was slight to almost perfect, and generally higher than interreader agreement. Mean sensitivity (84.0% and 92.0% for CS-VIBE and VIBE-DIXON, P = 0.500) and specificity (60.0% and 55.4% for CS-VIBE and VIBE-DIXON, P = 0.327) were comparable for both sequences.

CONCLUSIONS: Compressed-sensing volume-interpolated breath-hold examination allows an increase of the through-plane spatial resolution of ultrafast dynamic contrast-enhanced magnetic resonance imaging compared with TWIST at a comparable in-plane IQ. Morphological assessment of lesions using CS-VIBE is comparable to VIBE-DIXON, which takes 18 times longer. Consequently, CS-VIBE enables 3D evaluation of breast lesions in ultrafast breast MRI.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Vreemann S, Rodriguez-Ruiz A, Nickel D, et al. Compressed Sensing for Breast MRI: Resolving the Trade-Off Between Spatial and Temporal Resolution. Invest Radiol. 2017. PMID: 28463932. doi:10.1097/RLI.0000000000000384

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。