基于模型迭代重建的超低剂量胎儿CT:一项前瞻性先导研究

Ultra-Low-Dose Fetal CT With Model-Based Iterative Reconstruction: A Prospective Pilot Study.
PubMed 文献 · AJR Am J Roentgenol · 2017-05-02

本研究通过体模实验设定基于模型迭代重建(MBIR)的超低剂量胎儿CT扫描条件,并前瞻性评估其诊断效能与辐射剂量。结果显示超低剂量胎儿CT的容积CT剂量指数为0.5 mGy,胎儿辐射暴露为0.7 mSv,所得图像能准确进行骨骼发育不良的产前诊断,且与本院现行CT方法相比辐射暴露大幅降低而诊断效能未减。

要点

  • 超低剂量胎儿CT采用MBIR,容积CT剂量指数0.5 mGy,胎儿辐射暴露0.7 mSv。
  • 该方案图像能准确产前诊断骨骼发育不良,诊断效能未见降低。
  • 与本院现行CT方法相比,超低剂量胎儿CT大幅降低辐射暴露。

核心结论

  • 基于MBIR的超低剂量胎儿CT可大幅降低辐射暴露,同时不降低骨骼发育不良的产前诊断效能。

研究设计与样本

  • 前瞻性先导研究,使用32 cm CT剂量指数体模分析管电流与图像噪声关系以设定超低剂量扫描条件,并评估该条件下CT图像的诊断效能及辐射暴露。

诊断标准与关键数值

  • 自适应统计迭代重建中噪声与剂量平方根近似成反比,MBIR中噪声与剂量四次方根成反比;超低剂量胎儿CT的容积CT剂量指数为0.5 mGy,胎儿辐射暴露为0.7 mSv。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: Prenatal diagnosis of skeletal dysplasia by means of 3D skeletal CT examination is highly accurate. However, it carries a risk of fetal exposure to radiation. Model-based iterative reconstruction (MBIR) technology can reduce radiation exposure; however, to our knowledge, the lower limit of an optimal dose is currently unknown. The objectives of this study are to establish ultra-low-dose fetal CT as a method for prenatal diagnosis of skeletal dysplasia and to evaluate the appropriate radiation dose for ultra-low-dose fetal CT.

SUBJECTS AND METHODS: Relationships between tube current and image noise in adaptive statistical iterative reconstruction and MBIR were examined using a 32-cm CT dose index (CTDI) phantom. On the basis of the results of this examination and the recommended methods for the MBIR option and the known relationship between noise and tube current for filtered back projection, as represented by the expression SD = (milliamperes)-0.5, the lower limit of the optimal dose in ultra-low-dose fetal CT with MBIR was set. The diagnostic power of the CT images obtained using the aforementioned scanning conditions was evaluated, and the radiation exposure associated with ultra-low-dose fetal CT was compared with that noted in previous reports.

RESULTS: Noise increased in nearly inverse proportion to the square root of the dose in adaptive statistical iterative reconstruction and in inverse proportion to the fourth root of the dose in MBIR. Ultra-low-dose fetal CT was found to have a volume CTDI of 0.5 mGy. Prenatal diagnosis was accurately performed on the basis of ultra-low-dose fetal CT images that were obtained using this protocol. The level of fetal exposure to radiation was 0.7 mSv.

CONCLUSION: The use of ultra-low-dose fetal CT with MBIR led to a substantial reduction in radiation exposure, compared with the CT imaging method currently used at our institution, but it still enabled diagnosis of skeletal dysplasia without reducing diagnostic power.

来源:AJR Am J Roentgenol,查看原文。版权归原作者所有。