文献速递腹部技术与安全CT

新型探测器基光谱CT扫描仪生成的常规诊断腹部图像质量:体模与临床研究的技术报告

Quality of routine diagnostic abdominal images generated from a novel detector-based spectral CT scanner: a technical report on a phantom and clinical study.
PubMed 文献 · Abdom Radiol (NY) · 2017-01-01

本技术报告在体模和50例腹部CT扫描中评估新型探测器基光谱探测器CT(SDCT)生成的常规诊断图像质量,并与传统能量积分探测器CT(Brilliance iCT)比较。体模研究显示SDCT满足ACR对CT值、偏差、CNR和有效辐射剂量的要求;患者图像主观评分SDCT显著高于iCT,SNR在肾和主动脉显著更高,CNR在所有腹部器官均更高。结论认为SDCT常规诊断图像质量与传统CT相当,适合诊断用途。

要点

  • SDCT常规诊断图像在增强、噪声和整体质量的主观评分上显著高于iCT。
  • SDCT的CNR在所有腹部器官均高于iCT,SNR在肾脏和主动脉显著更高。
  • 体模研究中SDCT满足ACR对CT值、偏差、CNR和有效辐射剂量的要求。

核心结论

  • SDCT生成的常规诊断图像质量与传统能量积分探测器CT相当,适合诊断用途。

研究设计与样本

  • 体模研究依据ACR指南在SDCT和iCT上按成人腹部协议进行;回顾性分析各50例腹部CT扫描,在腹部多个位置放置相同ROI测量衰减、噪声、SNR和CNR,并由2名读者以5点Likert量表对增强、噪声和图像质量进行主观评分。

诊断标准与关键数值

  • 体模研究中SDCT图像满足ACR对CT数和偏差、CNR及有效辐射剂量的要求。
  • 主观评分SDCT增强4.79±0.38 vs iCT 4.60±0.51(p=0.005)。
  • 主观评分SDCT噪声4.63±0.42 vs iCT 4.29±0.50(p=0.000)。
  • 主观评分SDCT图像质量4.85±0.32 vs iCT 4.57±0.50(p=0.000)。
  • SNR SDCT肾脏11.1±6.3 vs iCT 8.7±5.0(p=0.033)、主动脉14.2±6.2 vs iCT 11.0±4.9(p=0.007)。

临床应用与阅片要点

  • SDCT常规诊断图像可用于腹部诊断,主观图像质量优于iCT。
  • SNR在肝脏、脾脏、胰腺和腰椎的组间差异未达统计学显著(p分别为0.662、0.152、0.303、0.937)。
  • SDCT的CNR在所有腹部器官均高于iCT。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To evaluate the image quality of routine diagnostic images generated from a novel detector-based spectral detector CT (SDCT) and compare it with CT images obtained from a conventional scanner with an energy-integrating detector (Brilliance iCT), Routine diagnostic (conventional/polyenergetic) images are non-material-specific images that resemble single-energy images obtained at the same radiation, METHODS: ACR guideline-based phantom evaluations were performed on both SDCT and iCT for CT adult body protocol. Retrospective analysis was performed on 50 abdominal CT scans from each scanner. Identical ROIs were placed at multiple locations in the abdomen and attenuation, noise, SNR, and CNR were measured. Subjective image quality analysis on a 5-point Likert scale was performed by 2 readers for enhancement, noise, and image quality.

RESULTS: On phantom studies, SDCT images met the ACR requirements for CT number and deviation, CNR and effective radiation dose. In patients, the qualitative scores were significantly higher for the SDCT than the iCT, including enhancement (4.79 ± 0.38 vs. 4.60 ± 0.51, p = 0.005), noise (4.63 ± 0.42 vs. 4.29 ± 0.50, p = 0.000), and quality (4.85 ± 0.32, vs. 4.57 ± 0.50, p = 0.000). The SNR was higher in SDCT than iCT for liver (7.4 ± 4.2 vs. 7.2 ± 5.3, p = 0.662), spleen (8.6 ± 4.1 vs. 7.4 ± 3.5, p = 0.152), kidney (11.1 ± 6.3 vs. 8.7 ± 5.0, p = 0.033), pancreas (6.90 ± 3.45 vs 6.11 ± 2.64, p = 0.303), aorta (14.2 ± 6.2 vs. 11.0 ± 4.9, p = 0.007), but was slightly lower in lumbar-vertebra (7.7 ± 4.2 vs. 7.8 ± 4.5, p = 0.937). The CNR of the SDCT was also higher than iCT for all abdominal organs.

CONCLUSION: Image quality of routine diagnostic images from the SDCT is comparable to images of a conventional CT scanner with energy-integrating detectors, making it suitable for diagnostic purposes.

来源:Abdom Radiol (NY),查看原文。版权归原作者所有。

引用

  1. Hojjati M, Van Hedent S, Rassouli N, et al. Quality of routine diagnostic abdominal images generated from a novel detector-based spectral CT scanner: a technical report on a phantom and clinical study. Abdom Radiol (NY). 2017. PMID: 28493070. doi:10.1007/s00261-017-1170-z

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。