径向超短TE成像使肝脏铁过载R2*磁共振定量无需屏气
研究纳入46例无法完成屏气动作的铁过载患者(38例镇静、8例未镇静)及16例可屏气对照,比较自由呼吸多回波GRE与径向超短TE成像。径向超短TE图像无运动伪影,平均R2*值与屏气多回波GRE无显著差异;自由呼吸多回波GRE因运动伪影R2*值显著偏低且变异系数约高两倍。作者认为径向超短TE是镇静儿童及无法屏气患者准确评估肝铁含量的可行替代方案。
要点
- 无法屏气患者行自由呼吸径向超短TE成像可获得无伪影R2*图,平均R2*值与屏气多回波GRE无显著差异。
- 自由呼吸多回波GRE图像有运动伪影,R2*值显著偏低,R2*变异系数约为屏气GRE和自由呼吸超短TE的两倍。
- 自由呼吸多回波GRE的R2*变异系数随R2*值升高而增大,提示运动伪影对高铁负荷者的R2*估算影响更强。
核心结论
- 自由呼吸多回波超短TE成像因径向采样对运动不敏感,图像质量好,在与多次平均自由呼吸多回波GRE相同的采集时间内获得准确R2*估算。
- 因此,自由呼吸多回波超短TE成像是镇静儿童及无法完成屏气动作患者准确评估肝铁含量的可行替代方案。
研究设计与样本
- 研究为影像学方法对照研究,纳入46例无法完成屏气动作的铁过载患者(38例镇静、8例未镇静)为试验组,16例可完成屏气动作的对照患者为对照组。
- 对照组行标准屏气多回波GRE、自由呼吸多回波GRE及自由呼吸多回波超短TE成像,试验组行自由呼吸多回波GRE及自由呼吸多回波超短TE成像。
- 计算定量R2*图,比较不同采集方法和患者组的平均肝脏R2*值及变异系数。
关键影像表现/征象
- 自由呼吸多回波GRE图像可见运动伪影,在对照组中其R2*值显著低于标准屏气多回波GRE和自由呼吸多回波超短TE,在试验组中显著低于自由呼吸多回波超短TE。
- 自由呼吸多回波超短TE图像产生的R2*图无伪影,平均R2*值与屏气多回波GRE所测值无显著差异。
- 自由呼吸多回波GRE图像的运动伪影使R2*变异系数约为屏气多回波GRE和自由呼吸多回波超短TE的R2*变异系数的两倍。
- 自由呼吸多回波超短TE的R2*变异系数在R2*值范围内相对恒定,而自由呼吸多回波GRE的R2*变异系数随R2*升高而增大,反映运动伪影随铁负荷增加对R2*估算影响更强。
诊断标准与关键数值
- 以平均肝脏R2*值及R2*变异系数作为定量比较指标,自由呼吸多回波超短TE的平均R2*值与屏气多回波GRE无显著差异。
- 自由呼吸多回波GRE的R2*变异系数约为屏气多回波GRE和自由呼吸多回波超短TE的R2*变异系数的两倍。
临床应用与阅片要点
- 对镇静儿童及无法完成屏气动作的患者,自由呼吸多回波超短TE成像可在相同采集时间内替代自由呼吸多回波GRE进行肝铁含量R2*定量。
- 自由呼吸多回波GRE图像出现运动伪影时R2*值偏低且变异系数增大,高铁负荷者受影响更明显,阅片和报告时需注意其R2*估算的可靠性。
- 作者认为自由呼吸多回波超短TE成像是此类患者准确评估肝铁含量的可行替代方案。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The objective of this study is to evaluate radial free-breathing (FB) multiecho ultrashort TE (UTE) imaging as an alternative to Cartesian FB multiecho gradient-recalled echo (GRE) imaging for quantitative assessment of hepatic iron content (HIC) in sedated patients and subjects unable to perform breath-hold (BH) maneuvers.
MATERIALS AND METHODS: FB multiecho GRE imaging and FB multiecho UTE imaging were conducted for 46 test group patients with iron overload who could not complete BH maneuvers (38 patients were sedated, and eight were not sedated) and 16 control patients who could complete BH maneuvers. Control patients also underwent standard BH multiecho GRE imaging. Quantitative R2* maps were calculated, and mean liver R2* values and coefficients of variation (CVs) for different acquisitions and patient groups were compared using statistical analysis.
RESULTS: FB multiecho GRE images displayed motion artifacts and significantly lower R2* values, compared with standard BH multiecho GRE images and FB multiecho UTE images in the control cohort and FB multiecho UTE images in the test cohort. In contrast, FB multiecho UTE images produced artifact-free R2* maps, and mean R2* values were not significantly different from those measured by BH multiecho GRE imaging. Motion artifacts on FB multiecho GRE images resulted in an R2* CV that was approximately twofold higher than the R2* CV from BH multiecho GRE imaging and FB multiecho UTE imaging. The R2* CV was relatively constant over the range of R2* values for FB multiecho UTE, but it increased with increases in R2* for FB multiecho GRE imaging, reflecting that motion artifacts had a stronger impact on R2* estimation with increasing iron burden.
CONCLUSION: FB multiecho UTE imaging was less motion sensitive because of radial sampling, produced excellent image quality, and yielded accurate R2* estimates within the same acquisition time used for multiaveraged FB multiecho GRE imaging. Thus, FB multiecho UTE imaging is a viable alternative for accurate HIC assessment in sedated children and patients who cannot complete BH maneuvers.