肝脏MR弹性成像的可重复性:一项Meta分析
该Meta分析系统检索2006至2015年文献,纳入12项研究共274例患者,估算肝脏MR弹性成像的可重复性系数(RC)为22%(95%CI:16.1%,28.2%),异质性主要来源为操作者是否受训、两次检查间隔时间及MR设备场强,为纵向定量影像生物标志物联盟(QIBA)的声明草案提供了依据。
要点
- 肝脏MR弹性成像的汇总可重复性系数为22%(95%CI:16.1%,28.2%)。
- 在相同部位、相同设备和采集序列下,肝硬度变化≥22%可认为真实变化(95%置信度)。
- 未受训操作者、1.5-T设备及较长检查间隔的研究RC估计值趋于更高。
核心结论
- Meta分析结果为纵向定量影像生物标志物联盟(QIBA)MR弹性成像声明草案提供依据:在相同部位、相同设备和采集序列下,肝硬度测量变化≥22%表明发生了真实的硬度变化(95%置信度)。
研究设计与样本
- Meta分析,系统检索2006至2015年发表的MR弹性成像文献,纳入12项研究共274例患者进行分析,所有纳入研究满足QUADAS-2七个类别中的四项或以上。
诊断标准与关键数值
- 汇总可重复性系数(RC)为22%(95%CI:16.1%,28.2%);肝硬度变化≥22%在相同部位、相同设备和采集序列下提示真实变化(95%置信度)。
局限性
- 异质性主要来源为受训与未受训操作者勾画轮廓选择感兴趣区、两次重复检查之间的时间以及MR成像设备的场强。
- 未使用训练有素操作者、使用1.5-T场强成像设备及检查间隔较长的研究,其RC估计值趋于更高。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Purpose To perform a meta-analysis to generate an estimate of the repeatability coefficient (RC) for magnetic resonance (MR) elastography of the liver. Materials and Methods A systematic search of databases was performed for publications on MR elastography during the 10-year period between 2006 and 2015. The identified studies were screened independently and were verified reciprocally by all authors. Two reviewers independently determined the percentage RC and effective sample size from each article. A forest plot was constructed of the percentage RC estimates from the 12 studies. Bootstrap 95% confidence intervals (CIs) were constructed for the summary percentage RCs. Results Twelve studies comprising 274 patients met the eligibility criteria and were included for analysis. A flow diagram of studies included according to Preferred Reporting Items for Systematic Reviews and Meta-Analyses guidelines was prepared for the inclusion and exclusion criteria. All studies included in the meta-analysis fulfilled four or more of the seven categories of the Quality Assessment of Diagnostic Accuracy Studies (QUADAS)-2. The estimated summary RC was 22% (95% CI: 16.1%, 28.2%). The three main sources for this heterogeneity were the trained versus untrained operator drawing contours to choose regions of interest, the time between two replicate examinations, and, finally, the field strength of the MR imaging unit. The RC estimates tended to be higher for studies that did not use a well-trained operator, those with 1.5-T field strength imaging units, and those with longer time intervals between examinations. Conclusion The meta-analysis results provide the basis for the following draft longitudinal Quantitative Imaging Biomarkers Alliance MR elastography claim: A measured change in hepatic stiffness of 22% or greater, at the same site and with use of the same equipment and acquisition sequence, indicates that a true change in stiffness has occurred with 95% confidence. © RSNA, 2017.
引用
- Serai SD, Obuchowski NA, Venkatesh SK, et al. Repeatability of MR Elastography of Liver: A Meta-Analysis. Radiology. 2017. PMID: 28530847. doi:10.1148/radiol.2017161398