CT增强比值检测血流动力学显著冠状动脉狭窄:猪模型验证研究

Detection of Hemodynamically Significant Coronary Artery Stenosis With CT Enhancement Ratio: A Validation Study in a Porcine Model.
PubMed 文献 · AJR Am J Roentgenol · 2017-05-15

在八只闭胸猪中模拟左前降支不同狭窄程度,腺苷负荷下行CT首过扫描,计算CT增强比值与CT衰减比值,并以微球法心肌血流量(MBF)比值和血流储备分数(FFR)为参考。CT增强比值与MBF比值强相关(r²=0.898),检测FFR定义显著缺血AUC为0.927,优于CT衰减比值(AUC=0.829)。

要点

  • CT增强比值与微球法MBF比值强相关(r²=0.898,p<0.001),优于CT衰减比值(r²=0.498)。
  • 以FFR定义显著缺血,CT增强比值AUC为0.927,CT衰减比值AUC为0.829。
  • CT增强比值可能作为心肌灌注相对定量评估的影像生物标志物。

核心结论

  • CT增强比值能可靠预测微球法测得的MBF比值,对血流动力学显著冠状动脉狭窄具有良好诊断性能。
  • CT增强比值可能作为心肌灌注相对定量评估的影像生物标志物。

研究设计与样本

  • 研究类型为猪模型验证研究,共八只闭胸猪。
  • 将球囊导管置入左前降支以模拟不同管腔狭窄程度。
  • 以彩色微球技术测量的MBF比值为参考,并以侵入性压力导丝测量FFR。
  • 在腺苷负荷试验期间行CT首过扫描,计算CT增强比值与CT衰减比值。

关键影像表现/征象

  • CT增强比值与微球法MBF比值强相关(y=0.07245+0.09963x;r²=0.898;p<0.001)。
  • CT衰减比值与微球法MBF比值仅中度相关(y=-1.5508+2.2684x;r²=0.498;p<0.001)。

诊断标准与关键数值

  • 以FFR定义腺苷负荷试验期间显著缺血时,CT增强比值检测显著缺血的AUC为0.927。
  • CT衰减比值检测FFR定义显著缺血的AUC为0.829。

临床应用与阅片要点

  • 作者指出CT增强比值可能作为心肌灌注相对定量评估的影像生物标志物。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: Although numerous techniques that are based on CT number analysis have been proposed, the assessment of hemodynamically significant coronary artery stenosis remains a great challenge. The purpose of this study is to validate use of the CT enhancement ratio in the detection of hemodynamically significant coronary artery stenosis in a porcine model.

MATERIALS AND METHODS: Experiments involving eight closed-chest swine were performed. A balloon catheter was placed into the left anterior descending coronary artery to simulate different degrees of luminal stenosis. The myocardial blood flow (MBF) ratio was measured using the colored microsphere technique. The fractional flow reserve was measured using an invasive pressure wire. CT scans were performed during the first-pass phase, while the pigs were undergoing adenosine stress tests. The CT enhancement ratio and the CT attenuation ratio were calculated using data from the CT images obtained.

RESULTS: Results suggested that the CT enhancement ratio had a strong correlation (y = 0.07245 + 0.09963x; r2 = 0.898; p < 0.001) with the MBF ratio measured using the microsphere technique, whereas only moderate correlation (y = -1.5508 + 2.2684x; r2 = 0.498; p < 0.001) was noted between the CT attenuation ratio and the MBF ratio measured using the microsphere technique. In ROC curve analysis, the AUC values of the CT enhancement ratio and the CT attenuation ratio were 0.927 and 0.829, respectively, with regard to the detection of significant ischemia during adenosine stress tests, as defined by the fractional flow reserve.

CONCLUSION: The CT enhancement ratio provides a reliable prediction of the MBF ratio measured using the microsphere technique, indicating that this metric has good diagnostic performance in the detection of hemodynamically significant coronary artery stenosis. The CT enhancement ratio may have potential for use as an imaging biomarker for the relative quantitative assessment of myocardial perfusion.

来源:AJR Am J Roentgenol,查看原文。版权归原作者所有。