文献速递外周血管心血管CT

单源双能量CT血管成像中的迭代重建:低与超低对比剂剂量方案的可行性

Iterative reconstruction in single-source dual-energy CT angiography: feasibility of low and ultra-low volume contrast medium protocols.
PubMed 文献 · Br J Radiol · 2017-06-23

本研究将120例已知或疑似主动脉夹层患者等分为3组行主动脉CTA,A组采用常规120kVp扫描、对比剂70ml、注射流率5ml/s,B组和C组采用单源双能量CT扫描,对比剂剂量分别为0.6和0.4ml/kg,并分别应用40%和50%自适应统计迭代重建。结果显示B组和C组的主观评分和CT值与A组相当,对比噪声比更高或相当,研究组平均有效辐射剂量较对照组降低22.5%-24.5%。结论认为联合应用自适应统计迭代重建和单源双能量CT,采用低和超低对比剂剂量与注射流率获取良好质量图像是可行的。

要点

  • B组和C组的主观评分和CT值与A组相当,对比噪声比更高或相当。
  • B组对比剂剂量和注射流率较A组分别降低40.1%和30%。
  • C组对比剂剂量和注射流率较B组分别进一步降低19.2%和22%。

核心结论

  • 联合应用自适应统计迭代重建和单源双能量CT行主动脉CTA时,采用低和超低对比剂剂量与注射流率获取良好质量图像是可行的。

研究设计与样本

  • 120例已知或疑似主动脉夹层患者等分为3组行主动脉CTA;A组为常规120kVp扫描、对比剂70ml、注射流率5ml/s,B组和C组为单源双能量CT扫描、对比剂剂量分别为0.6和0.4ml/kg、分别应用40%和50%自适应统计迭代重建。

诊断标准与关键数值

  • 采用五点分级方案主观评价图像质量。
  • 记录CT值和对比噪声比作为客观指标。
  • B组和C组的主观评分和CT值与A组相当,对比噪声比更高或相当。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To evaluate the feasibility of using contrast medium (CM) of low and ultra-low volumes and injection rates in aortic CT angiography (CTA) through the joint application of single-source dual-energy CT (ssDECT) and adaptive statistical iterative reconstruction (ASIR).

METHODS: 120 patients with known or suspected aortic dissection underwent aortic CTA and were equally divided into 3 groups. Conventional 120-kVp scan with a CM volume of 70 ml and an injection rate of 5 ml s-1 was performed on Group A. Groups B and C underwent ssDECT scan with CM volumes of 0.6 and 0.4 ml kg-1, respectively. 40% and 50% ASIR algorithms were applied for Groups B and C, respectively. A five-point grading scheme was utilized to subjectively evaluate the image quality, and the CT value and contrast-to-noise ratio were recorded as objective measures. The radiation dose was also evaluated.

RESULTS: Groups B and C had equivalent subjective scores and CT values as Group A, whereas they had higher or equivalent contrast-to-noise ratios. Group B had 40.1% and 30% reductions on CM volume and injection rate, respectively, than Group A. Group C further resulted in 19.2% and 22% lesser CM volume and injection rate than Group B. The average effective radiation doses for the study groups were 22.5-24.5% lower than the control group.

CONCLUSION: With the aid of ASIR and ssDECT for aortic CTA, it is feasible to adopt low and ultra-low CM volumes and injection rates while obtaining good quality images. Advances in knowledge: Low and ultra-low CM volumes and injection rates are feasible in CTA through the joint application of ssDECT and ASIR.

来源:Br J Radiol,查看原文。版权归原作者所有。

引用

  1. Hou P, Feng X, Liu J, et al. Iterative reconstruction in single-source dual-energy CT angiography: feasibility of low and ultra-low volume contrast medium protocols. Br J Radiol. 2017. PMID: 28555508. doi:10.1259/bjr.20160506

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。