不同厂商应变成像检测区域功能异常的准确性差异:EACVI-ASE应变标准化工作组报告
该研究用7台超声仪在5天内检查63名受试者(5名健康志愿者和58例既往心肌梗死患者),以心脏磁共振延迟钆增强为参考标准,比较6种厂商专用和2种独立应变分析软件检测透壁瘢痕节段的准确性。各厂商软件识别透壁瘢痕的AUC为0.74~0.83(收缩期峰值应变和收缩末期应变)及0.70~0.78(收缩后应变),厂商间差异有统计学意义(p<0.05),且所有厂商对基底段瘢痕的检出准确性均显著低于心尖段(p<0.05)。
要点
- 不同厂商应变软件识别区域功能异常的准确性存在显著差异(p<0.05)。
- 各厂商软件识别透壁瘢痕的AUC:收缩期峰值应变和收缩末期应变为0.74~0.83,收缩后应变为0.70~0.78。
- 所有厂商检测基底段瘢痕的准确性均显著低于心尖段(p<0.05)。
核心结论
- 不同厂商应变分析软件识别区域功能异常的准确性存在显著差异,且所有厂商对基底段瘢痕的检出准确性均显著低于心尖段。
研究设计与样本
- 研究纳入63名受试者(5名健康志愿者和58例既往心肌梗死患者),在5天内用7种不同超声仪检查,以心脏磁共振延迟钆增强为参考标准,采用6种厂商专用软件和2种独立应变分析软件测量节段收缩期峰值应变、收缩末期应变和收缩后应变。
诊断标准与关键数值
- 正常节段平均应变值范围:收缩期峰值应变-15.1%~-20.7%,收缩末期应变-14.9%~-20.6%,收缩后应变-16.1%~-21.4%;透壁瘢痕节段应变值范围:收缩期峰值应变-7.4%~-11.1%,收缩末期应变-7.7%~-10.8%,收缩后应变-10.5%~-14.3%;识别透壁瘢痕的AUC:收缩期峰值应变和收缩末期应变0.74~0.83,收缩后应变0.70~0.78。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: The purpose of this study was to compare the accuracy of vendor-specific and independent strain analysis tools to detect regional myocardial function abnormality in a clinical setting.
BACKGROUND: Speckle tracking echocardiography has been considered a promising tool for the quantitative assessment of regional myocardial function. However, the potential differences among speckle tracking software with regard to their accuracy in identifying regional abnormality has not been studied extensively.
METHODS: Sixty-three subjects (5 healthy volunteers and 58 patients) were examined with 7 different ultrasound machines during 5 days. All patients had experienced a previous myocardial infarction, which was characterized by cardiac magnetic resonance with late gadolinium enhancement. Segmental peak systolic (PS), end-systolic (ES) and post-systolic strain (PSS) measurements were obtained with 6 vendor-specific software tools and 2 independent strain analysis tools. Strain parameters were compared between fully scarred and scar-free segments. Receiver-operating characteristic curves testing the ability of strain parameters and derived indexes to discriminate between these segments were compared among vendors.
RESULTS: The average strain values calculated for normal segments ranged from -15.1% to -20.7% for PS, -14.9% to -20.6% for ES, and -16.1% to -21.4% for PSS. Significantly lower values of strain (p < 0.05) were found in segments with transmural scar by all vendors, with values ranging from -7.4% to -11.1% for PS, -7.7% to -10.8% for ES, and -10.5% to -14.3% for PSS. Accuracy in identifying transmural scar ranged from acceptable to excellent (area under the curve 0.74 to 0.83 for PS and ES and 0.70 to 0.78 for PSS). Significant differences were found among vendors (p < 0.05). All vendors had a significantly lower accuracy to detect scars in the basal segments compared with scars in the apex (p < 0.05).
CONCLUSIONS: The accuracy of identifying regional abnormality differs significantly among vendors.
引用
- Mirea O, Pagourelias ED, Duchenne J, et al. Intervendor Differences in the Accuracy of Detecting Regional Functional Abnormalities: A Report From the EACVI-ASE Strain Standardization Task Force. JACC Cardiovasc Imaging. 2017. PMID: 28528162. doi:10.1016/j.jcmg.2017.02.014
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。