胸腹部碳离子束扫描治疗中门控窗口对剂量分布与成像剂量的依赖性

Gating window dependency on scanned carbon-ion beam dose distribution and imaging dose for thoracoabdominal treatment.
PubMed 文献 · Br J Radiol · 2017-05-25

该回顾性研究利用10例肺癌和肝癌患者的4DCT图像制定碳离子扫描束治疗计划,评估不同门控窗口(占空比0–100%)下剂量适形度与透视成像剂量的权衡。肺癌病例中50%和30%占空比的V20肺分别为100%占空比的2.5%和6.0%,总成像剂量分别增加43.5%和115.8%;作者结论认为当正常组织剂量低于耐受水平时应扩大门控窗口以减少成像剂量和治疗时间。

要点

  • 门控窗口以肿瘤位置而非呼吸时相定义,占空比从0–100%在4DCT上变化。
  • 肺癌病例各门控窗口均达到D95-CTV>99%的剂量适形。
  • 50%和30%占空比的总成像剂量较100%占空比分别增加43.5%和115.8%。

核心结论

  • 当正常组织剂量低于耐受水平时,应扩大门控窗口以尽量减少成像剂量和治疗时间。

研究设计与样本

  • 回顾性研究,使用10例肺癌和肝癌患者的4DCT图像进行碳离子扫描束4D治疗计划,成像剂量由两个斜向方向的图像数乘以每幅图像空气比释动能计算,必要束流时间从治疗日志文件计算。

诊断标准与关键数值

  • 肺癌病例各门控窗口均达到D95-CTV>99%的剂量适形。
  • 肺癌病例50%和30%占空比的V20肺分别为100%占空比的2.5%和6.0%。
  • 肺癌病例30%占空比的脊髓/食管最大剂量较100%占空比分别降低6.8%/7.4%。
  • 肝癌病例50%和30%占空比的V10肝分别为100%占空比的9.4%和12.8%。
  • 肝癌病例50%和30%占空比的脊髓/食管最大剂量较100%占空比分别降低17.2%/19.3%和24.6%/29.8%。

临床应用与阅片要点

  • 作者结论认为,当正常组织剂量低于耐受水平时,应扩大门控窗口以尽量减少成像剂量和治疗时间。
  • 作者指出成像所致皮肤剂量可能无法抵消OAR剂量,而成像剂量是特别重要的因素。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To explore the trade-off between dose assessment and imaging dose in respiratory gating with radiographic fluoroscopic imaging, we evaluated the relationship between dose assessment and fluoroscopic imaging dose in various gating windows, retrospectively.

METHODS: Four-dimensional (4D) CT images acquired for 10 patients with lung and liver tumours were used for 4D treatment planning for scanned carbon ion beam. Imaging dose from two oblique directions was calculated by the number of images multiplied by the air kerma per image. Necessary beam-on time was calculated from the treatment log file. Accumulated dose distribution was calculated. The gating window was defined as tumour position not respiratory phase and changed from 0-100% duty cycle on 4DCT. These metrics were individually evaluated for every case.

RESULTS: For lung cases, sufficient dose conformation was achieved in respective gating windows [D95-clinical target volume (CTV) > 99%]. V20-lung values for 50%- and 30%-duty cycles were 2.5% and 6.0% of that for 100%-duty cycle. Maximum doses (cord/oesophagus) for 30%-duty cycle decreased 6.8%/7.4% to those for 100%-duty cycle. For liver cases, V10-liver values for 50%- and 30%-duty cycles were 9.4% and 12.8% of those for 100%-duty cycle, respectively. Maximum doses (cord/oesophagus) for 50%- and 30%-duty cycles also decreased 17.2%/19.3% and 24.6%/29.8% to those for 100%-duty cycle, respectively. Total imaging doses increased 43.5% and 115.8% for 50%- and 30%-duty cycles to that for the 100%-duty cycle.

CONCLUSION: When normal tissue doses are below the tolerance level, the gating window should be expanded to minimize imaging dose and treatment time. Advances in knowledge: The skin dose from imaging might not be counterbalanced to the OAR dose; however, imaging dose is a particularly important factor.