PET/MRI与PET/CT采集的FDG脑显像在癫痫灶检测中的诊断准确性比较:一项前瞻性研究
前瞻性纳入35例2-19岁定位相关癫痫患儿,比较PET/MR与PET/CT采集的FDG脑显像定位癫痫灶的诊断准确性,以多学科癫痫手术讨论确定的癫痫起源部位为参考标准。两种方式图像质量无显著差异,PET/MR准确性不劣于PET/CT(P=0.017)。
要点
- PET/MR采集的FDG脑显像定位癫痫灶准确性不劣于PET/CT(P=0.017)。
- 两种方式图像质量Likert评分无显著差异。
- 参考标准为多学科癫痫手术讨论确定的癫痫起源部位。
核心结论
- 经MR衰减校正的PET/MR采集FDG-PET图像诊断准确性不劣于传统CT衰减校正的PET图像。
研究设计与样本
- 前瞻性研究,35例临床转诊行FDG-PET/CT脑显像的定位相关癫痫患者,在同一PET/MR系统额外采集FDG-PET,五种衰减校正FDG图像由五位专家独立盲法判读,以多学科癫痫手术讨论确定的癫痫起源部位为参考标准,采用配对数据非劣性检验。
诊断标准与关键数值
- 采用Likert 1-5分评估图像质量,两种方式图像质量无显著差异;PET/MR定位癫痫灶准确性不劣于PET/CT(P=0.017)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: There is great interest in positron emission tomography (PET)/magnetic resonance (MR) as a clinical tool due to its capacity to provide diverse diagnostic information in a single exam.
OBJECTIVE: The goal of this exam is to compare the diagnostic accuracy of PET/MR-acquired [F-18]2-fluoro-2-deoxyglucose (FDG) brain exams to that of PET/CT with respect to identifying seizure foci in children with localization-related epilepsy.
MATERIALS AND METHODS: Institutional Review Board approval and informed consent were obtained for this Health Insurance Portability and Accountability Act-compliant, prospective study. All patients referred for clinical FDG-PET/CT exams of the brain at our institution for a diagnosis of localization-related epilepsy were prospectively recruited to undergo an additional FDG-PET acquisition on a tandem PET/MR system. Attenuation-corrected FDG images acquired at PET/MR and PET/CT were interpreted independently by five expert readers. Readers were blinded to the scanner used for acquisition and attenuation correction as well as all other clinical and imaging data. A Likert scale scoring system (1-5) was used to assess image quality. The locale of seizure origin determined at multidisciplinary epilepsy surgery work rounds was considered the reference standard. Non-inferiority testing for paired data was used to compare the diagnostic accuracy of PET/MR to that of PET/CT.
RESULTS: The final study population comprised 35 patients referred for a diagnosis of localization-related epilepsy (age range: 2-19 years; median: 11 years; 21 males, 14 females). Image quality did not differ significantly between the two modalities. The accuracy of PET/MR was not inferior to that of PET/CT for localization of a seizure focus (P=0.017).
CONCLUSION: The diagnostic accuracy of FDG-PET images acquired on a PET/MR scanner and generated using MR-based attenuation correction was not inferior to that of PET images processed by traditional CT-based correction.
引用
- Paldino MJ, Yang E, Jones JY, et al. Comparison of the diagnostic accuracy of PET/MRI to PET/CT-acquired FDG brain exams for seizure focus detection: a prospective study. Pediatr Radiol. 2017. PMID: 28512714. doi:10.1007/s00247-017-3888-8