文献速递神经外周血管MRDSA

无对比剂静音扫描MR血管成像评估低剖面可视腔内支撑装置治疗颅内前循环动脉瘤

Non-Contrast-Enhanced Silent Scan MR Angiography of Intracranial Anterior Circulation Aneurysms Treated with a Low-Profile Visualized Intraluminal Support Device.
PubMed 文献 · AJNR Am J Neuroradiol · 2017-05-18

本研究用静音MR血管成像(Silent MRA)、3D TOF-MRA和X线DSA评估31枚经低剖面可视腔内支撑装置(LVIS)支架治疗的颅内前循环动脉瘤。Silent MRA重建动脉显示质量评分3.16±0.79,显著高于3D TOF-MRA的1.48±0.67(P<.05);动脉瘤闭塞率DSA为69%/31%,Silent MRA为65%/35%,3D TOF-MRA为92%/8%。结论认为Silent MRA似乎有助于显示LVIS支架治疗的颅内前循环动脉瘤。

要点

  • Silent MRA显示LVIS支架重建动脉质量评分3.16±0.79,优于3D TOF-MRA的1.48±0.67(P<.05)。
  • 动脉瘤闭塞率:DSA 69%/31%,Silent MRA 65%/35%,3D TOF-MRA 92%/8%。
  • DSA与Silent MRA的模态间一致性好(κ=0.88),DSA与3D TOF-MRA一致性差(κ=0.30)。

核心结论

  • 静音MR血管成像似乎有助于显示经低剖面可视腔内支撑装置支架治疗的颅内前循环动脉瘤。

研究设计与样本

  • 研究纳入31枚经低剖面可视腔内支撑装置支架治疗的动脉瘤,采用Silent MRA、3D TOF-MRA和X线DSA进行评估,以4分法(1分不可见至4分优)对重建动脉MRA显示质量进行分级,以2级法(完全闭塞/残留[瘤颈或动脉瘤])评估动脉瘤闭塞状态,采用加权κ统计评估观察者间和模态间一致性。

诊断标准与关键数值

  • 重建动脉MRA显示质量采用4分法:1分不可见,4分优;动脉瘤闭塞状态采用2级法:完全闭塞/残留(瘤颈或动脉瘤);Silent MRA与3D TOF-MRA平均评分分别为3.16±0.79和1.48±0.67(P<.05),观察者间一致性κ=0.66;2级法闭塞率DSA为69%/31%,Silent MRA为65%/35%,3D TOF-MRA为92%/8%;模态间一致性DSA/Silent MRA为0.88,DSA/3D TOF-MRA为0.30。

临床应用与阅片要点

  • 作者认为Silent MRA似乎有助于显示经低剖面可视腔内支撑装置支架治疗的颅内前循环动脉瘤。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND AND PURPOSE: The Low-Profile Visualized Intraluminal Support Device comprises a small-cell nitinol structure and a single-wire braided stent that provides greater metal coverage than previously reported intracranial stents, as well as assumed strong susceptibility artifacts. This study aimed to assess the benefits of non-contrast-enhanced MRA by using a Silent Scan (Silent MRA) for intracranial anterior circulation aneurysms treated with Low-Profile Visualized Intraluminal Support Device stents.

MATERIALS AND METHODS: Thirty-one aneurysms treated with Low-Profile Visualized Intraluminal Support Device stents were assessed by using Silent MRA, 3D TOF-MRA, and x-ray DSA. The quality of MRA visualization of the reconstructed artery was graded on a 4-point scale from 1 (not visible) to 4 (excellent). Aneurysm occlusion status was evaluated by using a 2-grade scale (total occlusion/remnant [neck or aneurysm]). Weighted κ statistics were used to evaluate interobserver and intermodality agreement.

RESULTS: The mean scores ± SDs for Silent MRA and 3D TOF-MRA were 3.16 ± 0.79 and 1.48 ± 0.67 (P < .05), respectively, with substantial interobserver agreement (κ = 0.66). The aneurysm occlusion rates of the 2-grade scale (total occlusion/remnant [neck or aneurysm]) were 69%/31% for DSA, 65%/35% for Silent MRA, and 92%/8% for 3D TOF-MRA, respectively. The intermodality agreements were 0.88 and 0.30 for DSA/Silent MRA and DSA/3D TOF-MRA, respectively.

CONCLUSIONS: Silent MRA seems to be useful for visualizing intracranial anterior circulation aneurysms treated with Low-Profile Visualized Intraluminal Support Device stents.

来源:AJNR Am J Neuroradiol,查看原文。版权归原作者所有。

引用

  1. Takano N, Suzuki M, Irie R, et al. Non-Contrast-Enhanced Silent Scan MR Angiography of Intracranial Anterior Circulation Aneurysms Treated with a Low-Profile Visualized Intraluminal Support Device. AJNR Am J Neuroradiol. 2017. PMID: 28522664. doi:10.3174/ajnr.A5223

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。