肿瘤患者胸部、腹部及盆腔减剂量CT与标准剂量CT的头对头比较:基于模型的迭代重建对图像质量和病灶清晰度的影响

Reduced dose CT with model-based iterative reconstruction compared to standard dose CT of the chest, abdomen, and pelvis in oncology patients: intra-individual comparison study on image quality and lesion conspicuity.
PubMed 文献 · Abdom Radiol (NY) · 2017-01-01

前瞻性纳入44例肿瘤患者,在12个月内先后接受标准剂量(SD)CT分期和基于体重的减剂量(RD)CT联合基于模型的迭代重建(MBIR)。RD CT中位辐射剂量较SD CT降低46%(P<0.01),图像噪声更低、主观图像质量和病灶清晰度更优,靶病灶径线测量重复性高(SD CT ICC=0.987,RD CT ICC=0.97)。

要点

  • RD CT联合MBIR较SD CT中位剂量降低46%,图像噪声更低且主观图像质量更优。
  • RD CT病灶清晰度主观评分更优,但伪影更多,图像锐利度无显著差异(P=0.59)。
  • 靶病灶径线测量在SD CT(ICC=0.987)与RD CT(ICC=0.97)上均高度可重复。

核心结论

  • RD CT联合MBIR在肿瘤随访中可提供诊断级图像质量和与SD CT相当的病灶清晰度,同时中位剂量较SD CT降低46%。

研究设计与样本

  • 前瞻性自身对照研究,纳入44例12个月内曾接受SD CT分期的肿瘤患者,行基于体重的RD CT联合MBIR;记录辐射剂量,测量4种组织类型的衰减和图像噪声,分析每例最多5个靶病灶径线测量的可重复性,并由3名阅片者独立采用4分或5分Likert量表评估主观图像质量。

关键影像表现/征象

  • RD CT所有测量组织类型的中位图像噪声均低于SD CT(P<0.01)。
  • RD CT在图像噪声和整体图像质量方面的主观评分高于SD CT(P<0.01),图像锐利度无统计学显著差异(P=0.59)。
  • RD CT主观伪影多于SD CT(P<0.01),病灶清晰度主观评分优于SD CT(P<0.01)。

诊断标准与关键数值

  • 靶病灶径线重复测量可重复性:SD CT ICC=0.987,RD CT ICC=0.97。
  • 主观图像质量采用4分或5分Likert量表由3名阅片者独立评估。

临床应用与阅片要点

  • 肿瘤随访中采用基于体重的RD CT联合MBIR可在中位剂量降低46%的同时维持病灶清晰度和径线测量可重复性。
  • 阅片时需注意RD CT主观伪影较SD CT增多(P<0.01),但图像锐利度与SD CT无显著差异(P=0.59)。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To compare image quality and lesion conspicuity of reduced dose (RD) CT with model-based iterative reconstruction (MBIR) compared to standard dose (SD) CT in patients undergoing oncological follow-up imaging.

METHODS: Forty-four cancer patients who had a staging SD CT within 12 months were prospectively included to undergo a weight-based RD CT with MBIR. Radiation dose was recorded and tissue attenuation and image noise of four tissue types were measured. Reproducibility of target lesion size measurements of up to 5 target lesions per patient were analyzed. Subjective image quality was evaluated for three readers independently utilizing 4- or 5-point Likert scales.

RESULTS: Median radiation dose reduction was 46% using RD CT (P < 0.01). Median image noise across all measured tissue types was lower (P < 0.01) in RD CT. Subjective image quality for RD CT was higher (P < 0.01) in regard to image noise and overall image quality; however, there was no statistically significant difference regarding image sharpness (P = 0.59). There were subjectively more artifacts on RD CT (P < 0.01). Lesion conspicuity was subjectively better in RD CT (P < 0.01). Repeated target lesion size measurements were highly reproducible both on SD CT (ICC = 0.987) and RD CT (ICC = 0.97).

CONCLUSIONS: RD CT imaging with MBIR provides diagnostic imaging quality and comparable lesion conspicuity on follow-up exams while allowing dose reduction by a median of 46% compared to SD CT imaging.

来源:Abdom Radiol (NY),查看原文。版权归原作者所有。

引用

  1. Morimoto LN, Kamaya A, Boulay-Coletta I, et al. Reduced dose CT with model-based iterative reconstruction compared to standard dose CT of the chest, abdomen, and pelvis in oncology patients: intra-individual comparison study on image quality and lesion conspicuity. Abdom Radiol (NY). 2017. PMID: 28417170. doi:10.1007/s00261-017-1140-5