牙科及颌面部成像的锥形束CT:技术改进与低剂量协议

Cone beam computed tomography for dental and maxillofacial imaging: technique improvement and low-dose protocols.
PubMed 文献 · Radiol Med · 2017-04-01

本研究在Pax Zenith三维锥形束CT上评估五种不同采集协议的图像质量与辐射剂量,并用热释光剂量计在 anthropomorphic 体模58个敏感器官位点测量吸收剂量。大视野、常分辨率、80 kVp、5 mA、15秒协议的有效剂量为35微西弗,DAP从参考协议的1556 mGy cm²降至628 mGy cm²。作者认为该低剂量协议可用于需要完整评估牙科及颌面区域的精选病例以辅助治疗计划。

要点

  • 大视野、常分辨率、80 kVp、5 mA、15秒CBCT协议有效剂量为35微西弗。
  • kVp从95降至80时DAP从1556降至1013 mGy cm²,降低超过35%。
  • 高分辨率转常分辨率并将采集时间缩至15秒可再降剂量约40%至628 mGy cm²。

核心结论

  • 大视野、常分辨率、80 kVp、5 mA、15秒的CBCT低剂量协议有效剂量为35微西弗,可在精选病例中用于颌面区域完整评估以辅助治疗计划。

研究设计与样本

  • 在Pax Zenith三维锥形束CT(Vatech,韩国)及常规曲面体层和头颅成像设备上计算不同采集协议的剂量,使用 anthropomorphic 体模加载热释光剂量计在58个敏感器官位点测量吸收剂量,评估五种CBCT协议的图像质量与辐射剂量。

诊断标准与关键数值

  • 参考协议(大视野、高分辨率、95 kVp、5 mA、24秒)DAP为1556 mGy cm²;kVp从95降至80后DAP为1013 mGy cm²,降低超过35%;高分辨率转常分辨率并将采集时间缩至15秒后DAP为628 mGy cm²,进一步降低约40%,该协议有效剂量为35微西弗。

临床应用与阅片要点

  • 作者认为该低剂量协议可在需要完整评估牙科及颌面区域的精选病例中用于治疗计划。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: The aim of this study was to evaluate images quality and radiation doses of Cone Beam Computed Tomography (CBCT) for dental and maxillofacial imaging testing five different acquisition protocols.

METHODS: Dose measurements of different acquisition protocols were calculated for Pax Zenith three-dimensional (3D) Cone Beam (Vatech, Korea) and for conventional orthopantomography (OPT) and cephalometric skull imaging Ortophos (Sirona Dental Systems, Bernsheim, Germany). The absorbed organ doses were measured using an anthropomorphic phantom loaded with thermoluminescent dosimeters at 58 sites related to sensitive organs. Five different CBCT protocols were evaluated for image quality and radiation doses. They differed in FOV, image resolution, kVp, mA, acquisition time in seconds and radiation dose. Measurements were then carried out with the orthopantomograph. Equivalent and effective doses were calculated.

RESULTS: The reference protocol with large FOV, high resolution quality images, 95 kVp, 5 mA and acquisition time of 24 s resulted in a DAP value of 1556 mGy cm2 instead the protocol with reduced kVp from 95 to 80 kVp translated into a value of DAP inferior to 35% (from 1556 to 1013 mGy cm2). Going from a high resolution to a normal resolution, there was a reduction of the acquisition time to 15 s which allowed further dose reduction of approximately 40% (628 mGy cm2); this protocol resulted in a value of effective dose of 35 microSievert (μSv). Moreover, the effect of changing FOV has been evaluated, considering two scans with a reduced FOV (160 × 140 and 120 × 90 mm, respectively).

CONCLUSIONS: CBCT low-dose protocol with large FOV, normal resolution quality images, 80 kVp, 5 mA and acquisition time of 15 s resulted in a value of effective dose of 35 microSievert (μSv). This protocol allows the study of maxillofacial region with high quality of images and a very low radiation dose and, therefore, could be proposed in selected case where a complete assessment of dental and maxillofacial region is useful for treatment planning.

来源:Radiol Med,查看原文。版权归原作者所有。