3T MRI优化各向同性3D序列:无需延长采集时间即可准确描绘关节内病变
研究在3T MRI上优化各向同性3D TSE和GRE序列显示膝关节软骨病变,先用16名健康志愿者优化序列,再对57例膝关节软骨异常患者成像,共检出161个软骨病变(II级90个、III级71个)。TrueFISP检测软骨病变的敏感度和特异度最高,分别为84.7%和92%,但其软骨与半月板对比度最低。
要点
- 3D TrueFISP显示软骨优于3D SPACE序列。
- 3D TrueFISP检测低级别和高级别软骨缺损均可靠。
- 3T 3D TrueFISP提供软骨与关节液之间极佳对比。
核心结论
- 优化采集时间的各向同性3D TrueFISP在3T下能准确检测软骨缺损,但其软骨与半月板对比度最低。
研究设计与样本
- 研究论文;先用16名健康志愿者在3T MRI上优化实验性成像序列,再对57例膝关节软骨异常患者成像;序列包括3D PDW TSE(SPACE)含和不含脂肪抑制(FS)以及T2*W GRE(TrueFISP),采集时间分别为6分51秒、6分32秒和5分35秒。
关键影像表现/征象
- PDW SPACE-FS的软骨SNR均值最高,为72.2。
- TrueFISP的关节液/软骨CNR均值最高(101.5),关节液/韧带CNR均值最高(156.5),软骨/半月板CNR最低(48.5)。
- 三种序列在所有解剖结构上的SNR和CNR差异均有统计学意义(p值<0.05)。
诊断标准与关键数值
- 共检出161个确诊软骨病变,分级为II级90个、III级71个。
- TrueFISP检测软骨病变的敏感度和特异度最高,分别为84.7%和92%。
临床应用与阅片要点
- 3D TrueFISP在3T下可准确检测软骨缺损,阅片时需注意其软骨与半月板对比度最低。
- 3D TrueFISP检测低级别和高级别软骨缺损均可靠,软骨显示优于3D SPACE序列。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To investigate optimised isotropic 3D turbo spin echo (TSE) and gradient echo (GRE)-based pulse sequences for visualisation of articular cartilage lesions within the knee joint.
METHODS: Optimisation of experimental imaging sequences was completed using healthy volunteers (n=16) with a 3-Tesla (3T) MRI scanner. Imaging of patients with knee cartilage abnormalities (n=57) was then performed. Acquired sequences included 3D proton density-weighted (PDW) TSE (SPACE) with and without fat-suppression (FS), and T2*W GRE (TrueFISP) sequences, with acquisition times of 6:51, 6:32 and 5:35 min, respectively.
RESULTS: One hundred sixty-one confirmed cartilage lesions were detected and categorised (Grade II n=90, Grade III n=71). The highest sensitivity and specificity for detecting cartilage lesions were obtained with TrueFISP with values of 84.7% and 92%, respectively. Cartilage SNR mean for PDW SPACE-FS was the highest at 72.2. TrueFISP attained the highest CNR means for joint fluid/cartilage (101.5) and joint fluid/ligament (156.5), and the lowest CNR for cartilage/meniscus (48.5). Significant differences were identified across the three sequences for all anatomical structures with respect to SNR and CNR findings (p-value <0.05).
CONCLUSION: Isotropic TrueFISP at 3T, optimised for acquisition time, accurately detects cartilage defects, although it demonstrated the lowest contrast between cartilage and meniscus.
KEY POINTS: • Cartilage is better visualised with 3D TrueFISP than 3D SPACE sequences. • 3D TrueFISP is a reliable sequence for detecting low- and high-grade cartilage defects. • 3D TrueFISP at 3T provides excellent contrast between cartilage and joint fluid.
引用
- Abdulaal OM, Rainford L, MacMahon P, et al. 3T MRI of the knee with optimised isotropic 3D sequences: Accurate delineation of intra-articular pathology without prolonged acquisition times. Eur Radiol. 2017. PMID: 28432504. doi:10.1007/s00330-017-4816-x
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。