压缩弹性成像在跟腱中的重复性差:同日与连续日测量

Poor reproducibility of compression elastography in the Achilles tendon: same day and consecutive day measurements.
PubMed 文献 · Skeletal Radiol · 2017-04-04

研究纳入8名健康受试者,对跟腱行压缩弹性成像(CE)应变测量,评估同日连续5次及连续5天每日1次的重复性与不同足位下的可重复性。结果显示所有CV均高于53%,R值0.01-0.25,ICC均属差(0.00-0.11);长度与直径测量CV低、可靠性高。结论认为该特定系统CE测量跟腱硬度的可重复性低。

要点

  • 跟腱压缩弹性成像应变数据CV均高于53%,ICC 0.00-0.11属差。
  • 同日与连续日测量间R值0.01-0.25,呈无至弱相关。
  • 长度与直径测量CV低,提示常规形态测量可靠性高。

核心结论

  • 使用该特定系统对活体人跟腱硬度行压缩弹性成像,在同日连续测量、连续日测量及不同足位下可重复性低。

研究设计与样本

  • 研究纳入8名受试者(4男4女,平均年龄25.5±2.51岁,范围21-30岁;身高173.6±11.7 cm,范围156-189 cm),同日连续5次CE测量,并连续5天每日同一时间各1次,共80次测量,计算均值、标准差、变异系数(CV)、Pearson相关(R)及组内相关系数(ICC)。

诊断标准与关键数值

  • CE应变数据所有CV均高于53%。
  • CE测量间R值范围0.01-0.25,为无至弱相关。
  • CE的ICC值范围0.00-0.11,均归为差类别。
  • 长度与直径测量的CV可接受地低,提示可靠性高。

局限性

  • CE结果变异大,作者据此得出该特定系统测量活体人跟腱硬度在同日连续测量、连续日测量及不同足位下可重复性低的结论。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To determine the reproducibility of compression elastography (CE) when measuring strain data, a measure of stiffness of the human Achilles tendon in vivo, over consecutive measures, consecutive days and when using different foot positions.

MATERIALS AND METHODS: Eight participants (4 males, 4 females; mean age 25.5 ± 2.51 years, range 21-30 years; height 173.6 ± 11.7 cm, range 156-189 cm) had five consecutive CE measurements taken on one day and a further five CE measures taken, one per day, at the same time of day, every day for a consecutive 5-day period. These 80 measurements were used to assess both the repeatability and reproducibility of the technique. Means, standard deviations, coefficient of variation (CV), Pearson correlation analysis (R) and intra-class correlation coefficients (ICC) were calculated.

RESULTS: For CE data, all CVs were above 53%, R values indicated no-to-weak correlations between measures at best (range 0.01-0.25), and ICC values were all classified in the poor category (range 0.00-0.11). CVs for length and diameter measures were acceptably low indicating a high level of reliability.

CONCLUSIONS: Given the wide variation obtained in the CE results, it was concluded that CE using this specific system has a low level of reproducibility for measuring the stiffness of the human Achilles tendon in vivo over consecutive days, consecutive measures and in different foot positions.