水脂Dixon序列在乳房植入物评估中的应用:一种在临床实践中省时的快速方法提案
本研究回顾性评估27例3 T乳房MRI,比较单纯水脂Dixon序列与硅特异性联合水特异性序列对植入物完整性的诊断准确性;两种序列诊断准确率分别为100%和96.2%–100%,单纯Dixon序列可准确评估并缩短采集时间。
要点
- 单纯水脂Dixon序列(4分8秒)对植入物破裂诊断准确率96.2%–100%,与联合序列(7分17秒)相当。
- Dixon序列可出现硅胶与水之间的交换伪影(5例,18.5%),阅片时需注意。
- 联合序列水抑制均匀性中度-差的比例更高,运动伪影也更频繁。
核心结论
- 单一Dixon序列可对乳房植入物进行准确诊断评估,并同时缩短总采集时间。
研究设计与样本
- 经机构审查委员会(IRB)批准的回顾性研究,27例患者在3 T下行乳房MRI,两位阅片者独立定性评估两组序列,以手术确认或放射学报告诊断为标准参考。
关键影像表现/征象
- 两组序列中,植入物壁边缘界定、水抑制均匀性和整体图像质量在多数病例中评为良好-优秀。
- 组1(硅选择性水饱和STIR联合水-only Dixon)中水抑制均匀性中度-差的比例更高,运动伪影更频繁。
- 组2(脂-only和水-only Dixon)中5例(18.5%)出现硅胶与水之间的交换伪影。
- 两组序列在多数病例中诊断信心评为高-非常高。
诊断标准与关键数值
- 组1诊断准确率:阅片者1为100%,阅片者2为100%。
- 组2诊断准确率:阅片者1为96.2%,阅片者2为100%。
临床应用与阅片要点
- 单纯水脂Dixon序列(4分8秒)可用于乳房植入物评估,较联合序列(7分17秒)缩短采集时间。
- 阅片时注意Dixon序列可能出现硅胶与水交换伪影(5例,18.5%)。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
AIM: To evaluate the diagnostic accuracy achieved from a fat-water Dixon sequence alone compared to a combination of a silicone-specific magnetic resonance imaging (MRI) sequence and a water-specific MRI sequence in the assessment of breast implants.
MATERIALS AND METHODS: In this institutional review board (IRB)-approved study the integrity of breast implants was assessed retrospectively in 27 patients undergoing breast MRI at 3 T. A qualitative evaluation of (set 1) a silicon-selective water-saturated short tau inversion recovery (STIR) sequence in combination with a water-only Dixon dataset (total acquisition time 7 minutes 17 seconds), and of (set 2) fat-only and water-only Dixon datasets (4 minutes 8 seconds) was performed by two readers independently evaluating the following features: margin definition of the implant, water suppression homogeneity, image quality, presence of artefacts and their effects on the imaging interpretation, and diagnostic confidence. Diagnostic accuracy in implant rupture detection was determined and either surgical confirmation or diagnosis from the radiological report was used as a standard of reference.
RESULTS: In both sequences, margin definition of the implant wall, water suppression homogeneity, and overall image quality were rated good-excellent in most of cases. Water suppression homogeneity was moderate-poor in a greater number of cases in set 1. Movement artefacts were more frequent in set 1 whereas five cases (18.5%) exhibited swap artefacts between silicone and water in set 2. Diagnostic confidence was rated high-very high with both sequences in most of cases. Diagnostic accuracy was 100% for both readers using set 1 and 96.2% and 100% using set 2.
CONCLUSION: A single Dixon sequence allows an accurate diagnostic evaluation of breast implants and concomitant shortening of the overall acquisition time.
引用
- Marcon M, Becker AS, Ulbrich EJ, et al. Water-fat Dixon sequences in the evaluation of breast implants: proposal of a time effective rapid approach in the clinical practice. Clin Radiol. 2017. PMID: 28438322. doi:10.1016/j.crad.2017.03.020
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。