使用宽孔径3T MRI测量乳腺脂肪体积:传统乳腺X线密度评估与自动分割MRI密度测量的比较
研究纳入40例同时接受乳腺X线和动态乳腺MRI的女性,用两点Dixon技术、相位敏感重建和图谱分割获得脂肪-水分离图像,比较AMRA软件自动测量的乳腺密度与放射科医生BI-RADS密度分级。结果显示MRI自动密度与BI-RADS呈高度相关(左侧rho=0.73,右侧rho=0.75),且增强前后密度测量几乎完全一致(左侧rho=0.98、右侧rho=0.99);作者认为AMRA序列可全自动、可重复、客观评估纤维腺体组织比例,与乳腺X线密度评估相关性好且无电离辐射。
要点
- MRI自动分割测得的乳腺密度与BI-RADS密度分级相关性为左侧rho=0.73、右侧rho=0.75。
- 增强前后MRI乳腺密度测量几乎完全一致,左侧rho=0.98、右侧rho=0.99。
- BI-RADS密度分级的观察者间信度为加权Kappa k=0.8(95%CI 0.74-0.87)。
核心结论
- AMRA序列可全自动、可重复、客观地评估纤维腺体组织比例,与乳腺X线密度评估相关性好,并具有避免电离辐射的优势。
研究设计与样本
- 研究纳入40例接受乳腺X线和动态乳腺MRI的女性,采用两点Dixon技术、相位敏感重建和图谱分割获得脂肪-水分离图像,比较AMRA软件自动密度测量与BI-RADS四分类密度评估。
关键影像表现/征象
- MRI脂肪-水分离图像经图谱分割可自动得出乳腺腺体组织比例。
- 平扫乳腺密度均值左侧0.31±0.22、右侧0.29±0.21。
- 增强后乳腺密度均值左侧0.32±0.19、右侧0.32±0.2。
诊断标准与关键数值
- 平扫与增强后乳腺密度定量相关性:左侧Spearman rho=0.98(95%CI 0.97-0.99),右侧rho=0.99(95%CI 0.98-0.99)。
- 平扫与增强后密度测量的一致性界限:左侧-0.11至0.08,右侧-0.08至0.03。
- BI-RADS与MRI乳腺密度的Spearman相关系数:左侧rho=0.73(95%CI 0.60-0.82),右侧rho=0.75(95%CI 0.63-0.83)。
- BI-RADS密度分级的观察者间信度:加权Kappa k=0.8(95%CI 0.74-0.87)。
临床应用与阅片要点
- 作者认为AMRA序列可全自动、可重复、客观地评估纤维腺体组织比例。
- 作者指出该方法与乳腺X线密度评估相关性好,并具有避免电离辐射的优势。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
AIM: To compare magnetic resonance imaging (MRI)-derived breast density measurements using automatic segmentation algorithms with radiologist estimations using the Breast Imaging Reporting and Data Systems (BI-RADS) density classification.
MATERIALS AND METHODS: Forty women undergoing mammography and dynamic breast MRI as part of their clinical management were recruited. Fat-water separated MRI images derived from a two-point Dixon technique, phase-sensitive reconstruction, and atlas-based segmentation were obtained before and after intravenous contrast medium administration. Breast density was assessed using software from Advanced MR Analytics (AMRA), Linköping, Sweden, with results compared to the widely used four-quartile quantitative BI-RADS scale.
RESULTS: The proportion of glandular tissue in the breast on MRI was derived from the AMRA sequence. The mean unenhanced breast density was 0.31±0.22 (mean±SD; left) and 0.29±0.21 (right). Mean breast density on post-contrast images was 0.32±0.19 (left) and 0.32±0.2 (right). There was "almost perfect" correlation between pre- and post-contrast breast density quantification: Spearman's correlation rho=0.98 (95% confidence intervals [CI]: 0.97-0.99; left) and rho=0.99 (95% CI: 0.98-0.99; right). The 95% limits of agreement were -0.11-0.08 (left) and -0.08-0.03 (right). Interobserver reliability for BI-RADS was "substantial": weighted Kappa k=0.8 (95% CI: 0.74-0.87). The Spearman correlation coefficient between BI-RADS and MRI breast density was rho=0.73 (95% CI: 0.60-0.82; left) and rho=0.75 (95% CI: 0.63-0.83; right) which was also "substantial".
CONCLUSION: The AMRA sequence provides a fully automated, reproducible, objective assessment of fibroglandular breast tissue proportion that correlates well with mammographic assessment of breast density with the added advantage of avoidance of ionising radiation.
引用
- Petridou E, Kibiro M, Gladwell C, et al. Breast fat volume measurement using wide-bore 3 T MRI: comparison of traditional mammographic density evaluation with MRI density measurements using automatic segmentation. Clin Radiol. 2017. PMID: 28363661. doi:10.1016/j.crad.2017.02.014
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。