空间复合成像技术获取旋转弹性图:一项可行性研究

Spatial Compounding Technique to Obtain Rotation Elastogram: A Feasibility Study.
PubMed 文献 · Ultrasound Med Biol · 2017-04-19

本研究通过 Field II 仿真与组织仿体实验,探索空间复合成像能否改善旋转弹性图质量。仿体顶部接受约1%-2%轴向压缩,用多级二维位移追踪算法在不同入射角估计轴向与横向位移,以位移梯度差之半计算旋转弹性图,并用对比噪声比量化质量。仿真与实验结果一致,提示空间复合技术可显著改善旋转弹性图质量。

要点

  • 旋转弹性图通过轴向位移横向梯度与横向位移轴向梯度之差的一半计算获得。
  • 空间复合成像可改善旋转弹性图的对比噪声比,仿真与仿体实验结论一致。
  • 最大转向角与增量角会影响旋转弹性图质量,原文通过变化这两个参数评估其影响。

核心结论

  • 仿真与实验结果相互印证,提示使用空间复合成像技术可显著改善旋转弹性图的质量。

研究设计与样本

  • 可行性研究,采用 Field II 仿真与组织仿体实验;仿体顶部施加约1%-2%轴向压缩,用多级二维位移追踪算法在不同入射角获取角度轴向与横向位移估计。

诊断标准与关键数值

  • 旋转弹性图取轴向位移估计的横向梯度与横向位移估计的轴向梯度之差的一半;以对比噪声比量化旋转弹性图质量,并通过改变最大转向角与增量角评估其对旋转弹性图质量的影响。

局限性

  • 超声成像固有的横向分辨率较差,限制了局部刚体旋转的估计。
  • 本研究仅为仿真与组织仿体实验的可行性探索。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

The perception of stiffness and slipperiness of a breast mass on palpation is used by physicians to assess the level of suspicion of a lesion as being malignant or benign. However, most current ultrasound elastography imaging methods provide only stiffness-related information. There is no existing approach that provides information about the local rigid body rotation undergone by only a loosely bonded, asymmetrically oriented lesion subjected to a small quasi-static compression. The inherent poor lateral resolution in ultrasound imaging poses a limitation in estimating the local rigid body rotation. Several techniques have been reported in the literature to improve the lateral resolution in ultrasound imaging, and among them is spatial compounding. In this study, we explore the feasibility of obtaining better-quality rotation elastograms with spatial compounding through simulations using Field II and experiments on tissue-mimicking phantoms. The phantom was subjected to axial compression (∼1%-2%) from the top, and the angular axial and lateral displacement estimates were obtained using a multilevel 2-D displacement tracking algorithm at different insonification angles. A rotation elastogram (RE) was obtained by taking half of the difference between the lateral gradient of the axial displacement estimates and the axial gradient of the lateral displacement estimates. Contrast-to-noise ratio was used to quantify the improvements in quality of RE. Contrast-to-noise ratio values were calculated by varying the maximum steering angle and the incremental angle, and its effects on RE quality were evaluated. Both simulation and experimental results corroborated and indicated a significant improvement in the quality of RE using compounding technique.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Kothawala A, Chandramoorthi S, Reddy NRK, et al. Spatial Compounding Technique to Obtain Rotation Elastogram: A Feasibility Study. Ultrasound Med Biol. 2017. PMID: 28433440. doi:10.1016/j.ultrasmedbio.2017.01.026

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。