超越导丝与粒子:反射器引导的乳腺病灶定位与切除
单机构回顾性分析100例女性111个乳腺靶点、123枚Savi Scout反射器的定位与切除,超声引导占32.5%、钼靶引导占67.5%,全部靶点与反射器均被切除,切缘阳性需再切除率7.4%,无相关术后并发症。
要点
- Savi Scout反射器可用于乳腺病灶术前定位,本组123枚反射器全部置入并全部切除。
- 靶点与反射器平均钼靶距离0.3 cm,近距离(最近2.6 cm)可放置多枚反射器进行夹持定位。
- 54例恶性中4例(7.4%)切缘阳性需再切除,110例标本X线中5例(4.5%)靶点与反射器距离较术后当日钼靶增大超过1.0 cm。
核心结论
- 作者认为Savi Scout是定位和切除乳腺病灶的准确、可靠方法,切缘阳性率和再切除率可接受,并认为其克服了其他定位方法的许多局限、值得进一步研究。
研究设计与样本
- 单机构回顾性研究,纳入2015年6月至2016年5月接受Savi Scout手术导航系统乳腺病灶定位与切除的100例女性。
关键影像表现/征象
- 123枚无放射性、红外激活的电磁波反射器经皮置入111个乳腺靶点邻近或内部,其中20例患者放置2或3枚反射器用于夹持或多病灶定位,同侧时最近相距2.6 cm。
- 靶点与反射器距离在钼靶和标本X线上测量,反射器深度在钼靶上测量;靶点与反射器平均钼靶距离0.3 cm。
- 110份标本X线中5份(4.5%;95% CI:1.7%,10.4%)显示靶点与反射器距离较置入当日术后钼靶增大超过1.0 cm(范围1.1-2.6 cm),其中3份伴血肿、2份无明确原因迁移。
诊断标准与关键数值
- 123枚反射器全部置入(100%;95% CI:96.4%,100%),其中超声引导40枚(32.5%;95% CI:24.9%,41.2%)、钼靶引导83枚(67.5%;95% CI:58.8%,75.1%)。
- 123个靶点和反射器全部切除(100%;95% CI:96.4%,100%)。
- 病理示110个病灶中恶性54个(49.1%;95% CI:39.9%,58.3%;平均1.0 cm;范围0.1-5 cm)、高危病变32个(29.1%;95% CI:21.4%,38.2%)、良性24个(21.8%;95% CI:115.1%,30.4%,原文此处写作115.1%,按上下文应为15.1%)。
- 54例恶性中4例(7.4%;95% CI:2.4%,18.1%)切缘阳性需再切除。
- 未发现相关术后并发症。
临床应用与阅片要点
- 术前0-8天在影像引导下将反射器经皮置于靶点邻近或内部,术中用手持探头发出红外光和电磁波定位靶点与反射器。
- 标本X线及病理分析用于核实靶点和反射器是否切除,并可在钼靶与标本X线上测量靶点至反射器距离、在钼靶上测量反射器深度。
- 同侧多枚反射器可用于夹持或多病灶定位,最近可相距2.6 cm。
局限性
- 原文为单机构回顾性研究。
- 作者在结论中提出Savi Scout克服了其他定位方法的许多局限,值得进一步研究。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Purpose To evaluate outcomes of Savi Scout (Cianna Medical, Aliso Viejo, Calif) reflector-guided localization and excision of breast lesions by analyzing reflector placement, localization, and removal, along with target excision and rates of repeat excision (referred to as re-excision). Materials and Methods A single-institution retrospective review of 100 women who underwent breast lesion localization and excision by using the Savi Scout surgical guidance system from June 2015 to May 2016 was performed. By using image guidance 0-8 days before surgery, 123 nonradioactive, infrared-activated, electromagnetic wave reflectors were percutaneously inserted adjacent to or within 111 breast targets. Twenty patients had two or three reflectors placed for bracketing or for localizing multiple lesions, and when ipsilateral, they were placed as close as 2.6 cm apart. Target and reflector were localized intraoperatively by one of two breast surgeons who used a handpiece that emitted infrared light and electromagnetic waves. Radiographs of the specimen and pathologic analysis helped verify target and reflector removal. Target to reflector distance was measured on the mammogram and radiograph of the specimen, and reflector depth was measured on the mammogram. Pathologic analysis was reviewed. Re-excision rates and complications were recorded. By using statistics software, descriptive statistics were generated with 95% confidence intervals (CIs) calculated. Results By using sonographic (40 of 123; 32.5%; 95% CI: 24.9%, 41.2%) or mammographic (83 of 123; 67.5%; 95% CI: 58.8% 75.1%) guidance, 123 (100%; 95% CI: 96.4%, 100%) reflectors were placed. Mean mammographic target to reflector distance was 0.3 cm. All 123 (100%; 95% CI: 96.4%, 100%) targets and reflectors were excised. Pathologic analysis yielded 54 of 110 malignancies (49.1%; 95% CI: 39.9%, 58.3%; average, 1.0 cm; range, 0.1-5 cm), 32 high-risk lesions (29.1%; 95% CI: 21.4%, 38.2%), and 24 benign lesions (21.8%; 95% CI: 115.1%, 30.4%). Four of 54 malignant cases (7.4%; 95% CI: 2.4%, 18.1%) demonstrated margins positive for cancer that required re-excision. Five of 110 radiographs of the specimen (4.5%; 95% CI: 1.7%, 10.4%) demonstrated increased distance between the target and reflector distance of greater than 1.0 cm (range, 1.1-2.6 cm) compared with postprocedure mammogram the day of placement, three of five were associated with hematomas, two of five migrated without identifiable cause. No related postoperative complications were identified. Conclusion Savi Scout is an accurate, reliable method to localize and excise breast lesions with acceptable margin positivity and re-excision rates. Bracketing is possible with reflectors as close as 2.6 cm. Savi Scout overcomes many limitations of other localization methods, which warrants further study. © RSNA, 2017.