文献速递腹部技术与安全CT

腹部CT成像中70-keV虚拟单能谱图像评估:与传统120-kVp多能谱图像的比较研究

Assessment of 70-keV virtual monoenergetic spectral images in abdominal CT imaging: A comparison study to conventional polychromatic 120-kVp images.
PubMed 文献 · Abdom Radiol (NY) · 2017-01-01

前瞻性研究纳入55例患者,在光谱探测器CT门静脉期单次增强扫描后生成120-kVp多能谱与70-keV虚拟单能谱数据集。70-keV虚拟单能谱图像噪声显著更低、SNR与CNR显著更高,主观评分中位数4.0对3.0,整体图像质量读者间一致性良好。

要点

  • 70-keV虚拟单能谱腹部CT图像噪声显著低于120-kVp多能谱图像(p<0.001)。
  • 70-keV虚拟单能谱图像SNR与CNR均显著高于多能谱图像(p均<0.001)。
  • 主观图像质量各参数中位数:多能谱3.0 vs 虚拟单能谱4.0,虚拟单能谱显著更优(p<0.001)。

核心结论

  • 在腹部成像中,70-keV虚拟单能谱CT图像在噪声、SNR、CNR及主观评分方面均显著优于传统120-kVp多能谱图像。

研究设计与样本

  • 前瞻性研究,纳入55例患者(年龄54.3±16.8岁;男28例、女27例),使用光谱探测器CT(Philips Healthcare)门静脉期单次增强扫描后生成120-kVp多能谱与70-keV虚拟单能谱数据集。

关键影像表现/征象

  • 目标器官噪声在虚拟单能谱图像中显著低于多能谱图像(p<0.001)。
  • 虚拟单能谱图像SNR与CNR均显著高于多能谱图像(p均<0.001)。
  • 主观图像质量所有评估参数在70-keV虚拟单能谱图像中显著更优(p<0.001)。
  • 主观参数中位数评分:多能谱图像3.0 vs 虚拟单能谱图像4.0。
  • 整体图像质量读者间一致性良好:多能谱图像Kappa 0.767,虚拟单能谱图像Kappa 0.762。

诊断标准与关键数值

  • 主观图像质量采用5分Likert量表(1分为最差,5分为最佳),由两名放射科医师独立评分。
  • 测量指标包括肝脏、脾脏、胰腺、肾脏、主动脉、门静脉及肌肉的CT值(HU)、噪声、SNR与CNR。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To evaluate the image quality of 70-keV virtual monoenergetic (monoE) abdominal CT images compared to 120-kVp polychromatic images generated from a spectral detector CT (SDCT) scanner.

METHODS: This prospective study included generation of a 120-kVp polychromatic dataset and a 70-keV virtual monoE dataset after a single contrast-enhanced CT acquisition on a SDCT scanner (Philips Healthcare) during portal venous phase. The attenuation values (HU), noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) were measured in the liver, spleen, pancreas, kidney, aorta, portal vein, and muscle. The subjective image quality including noise, soft tissue contrast, sharpness, and overall image quality were graded on a 5-point Likert scale by two radiologists independently (1-worst image quality, 5-best image quality). Statistical analysis was performed using paired sample t test and Fleiss's Kappa.

RESULTS: Fifty-five patients (54.3 ± 16.8 y/o; 28 M, 27 F) were recruited. The noise of target organs was significantly lower in virtual monoE images in comparison to polychromatic images (p < 0.001). The SNR and CNR were significantly higher in virtual monoE images (p < 0.001 for both). Subjective image quality of 70-keV virtual monoE images was significantly better (p < 0.001) for all evaluated parameters. Median scores for all subjective parameters were 3.0 versus 4.0 for polychromatic vs virtual monoE images, respectively. The inter-reader agreement for overall image quality was good (Kappa were 0.767 and 0.762 for polychromatic and virtual monoE images, respectively).

CONCLUSION: In abdominal imaging, 70-keV virtual monoE CT images demonstrated significantly better noise, SNR, CNR, and subjective score compared to conventional 120-kVp polychromatic images.

来源:Abdom Radiol (NY),查看原文。版权归原作者所有。

引用

  1. Rassouli N, Chalian H, Rajiah P, et al. Assessment of 70-keV virtual monoenergetic spectral images in abdominal CT imaging: A comparison study to conventional polychromatic 120-kVp images. Abdom Radiol (NY). 2017. PMID: 28421243. doi:10.1007/s00261-017-1151-2

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。