短扫描长度腹部定量CT的辐射剂量评估
研究分析165例男性与233例女性共398次腹部定量CT检查,CTDIvol为4.66 mGy、扫描长度1 cm,比较1 cm扫描中心中点剂量DL(0)与CTDIvol及SSDE。结果显示平均SSDE为6.22±1.36 mGy、平均DL(0)为2.97±0.95 mGy,随患者直径增大DL(0)/CTDIvol比值从1.24降至0.25、DL(0)/SSDE比值从0.61降至0.38,提示短扫描长度剂量评估需同时考虑患者体型与扫描长度。
要点
- 1 cm短扫描长度下SSDE(平均6.22±1.36 mGy)明显高于实际中心中点剂量DL(0)(平均2.97±0.95 mGy)。
- 随患者直径增大,DL(0)/CTDIvol比值从1.24降至0.25,DL(0)/SSDE比值从0.61降至0.38。
- 短扫描长度剂量评估需同时考虑患者体型与扫描长度,遗漏任一因素可导致错误结果。
核心结论
- 腹部定量CT的患者剂量可能与CTDIvol和SSDE有较大差异,短扫描长度扫描的剂量评估必须同时考虑患者体型和扫描长度。
研究设计与样本
- 研究分析165例男性和233例女性共398次腹部定量CT检查,所有检查CTDIvol为4.66 mGy、扫描长度为1 cm。
诊断标准与关键数值
- 平均患者直径为28.5±6.3 cm(范围16.5-46.6 cm);平均SSDE为6.22±1.36 mGy(范围3.12-9.42 mGy);平均DL(0)为2.97±0.95 mGy(范围1.18-5.77 mGy)。
临床应用与阅片要点
- 研究提示短扫描长度(1 cm)的腹部定量CT剂量评估不能直接采用CTDIvol或SSDE,需同时考虑患者体型与扫描长度。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To assess radiation dose for patients who received abdominal quantitative CT and to compare the midpoint dose [DL(0)] at the centre of a 1-cm scan length with the volume CT dose index (CTDIvol). Although the size-specific dose estimate (SSDE) proposed in The American Association of Physicists in Medicine Report No. 204 is not applicable for short-length scans, commercial dose-monitoring software, such as Radimetrics™ Enterprise Platform (Bayer HealthCare, Whippany, NJ), reports SSDE for all scans. SSDE was herein compared with DL(0).
METHODS: Data were analyzed from 398 abdominal quantitative CT examinations in 165 males and 233 females. The CTDIvol was 4.66 mGy, and the scan length was 1 cm for all examinations. Radimetrics was used to extract patient diameter and SSDE. DL(0) was assessed using a previously reported method that takes into account both patient size and scan length.
RESULTS: The mean patient diameter was 28.5 ± 6.3 cm (range, 16.5-46.6 cm); the mean SSDE was 6.22 ± 1.36 mGy (range, 3.12-9.42 mGy); and the mean DL(0) was 2.97 ± 0.95 mGy (range, 1.18-5.77 mGy). As patient diameter increased, the DL(0) to CTDIvol ratio decreased, ranging from 1.24 to 0.25; the DL(0) to SSDE ratio also decreased, ranging from 0.61 to 0.38.
CONCLUSION: The dose to the patients from abdominal quantitative CT may be largely different from CTDIvol and SSDE. This study demonstrates the necessity of taking into account not only patient size but also scan length for evaluating the dose from short-length scans. Advances in knowledge: In CT examinations with 1-cm scan length, dose evaluation needs to take into account both patient size and scan length. An omission of either factor can result in an erroneous result.