文献速递腹部儿科影像MR

MR弹性成像:在儿童和年轻成人患者中技术成功率高

MR elastography: high rate of technical success in pediatric and young adult patients.
PubMed 文献 · Pediatr Radiol · 2017-04-03

本研究回顾性分析2011年9月至2015年8月间372例≤18岁患者共468次肝脏2-D GRE MR弹性成像检查,技术成功率为96%(450/468),失败原因包括桨板定位不佳致肝剪切波不足、铁过载、患者无法耐受MRI、呼吸/运动伪影、植入硬件伪影及技术故障,9次重复检查中7次(78%)成功。

要点

  • 儿童肝脏2-D GRE MR弹性成像在1.5 T下技术成功率达96%(450/468)。
  • 检查失败少见,主要与患者自身因素有关,部分可通过精细操作技术减轻。
  • 失败原因包括桨板定位不佳、铁过载、无法耐受MRI、呼吸/运动、植入物伪影和技术故障。

核心结论

  • 儿童1.5 T肝脏2-D GRE MR弹性成像技术稳健,检查失败不常见且主要反映患者特定因素,其中部分可通过精细操作技术减轻。

研究设计与样本

  • 回顾性影像记录检索,纳入2011年9月至2015年8月间≤18岁患者,在两台1.5-T MRI平台之一上行肝脏2-D GRE MR弹性成像,共468次检查(372例患者),复习影像报告和临床记录以确定失败采集、失败原因及是否随后成功检查。

诊断标准与关键数值

  • 技术成功率96%(450/468)。
  • 检查成功与失败患者平均年龄无显著差异(12.6±3.6岁 vs 11.2±4.1岁,P=0.12)。
  • 检查成功与失败患者BMI无显著差异(28.2±12.4 kg/m² vs 29.5±10 kg/m²,P=0.6)。
  • 9次重复检查中7次成功(78%)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Magnetic resonance (MR) elastography allows the noninvasive assessment of liver stiffness, which is a surrogate for fibrosis.

OBJECTIVE: The purpose of this study was to describe our experience using liver MR elastography in a large pediatric population with attention to the frequency and causes of exam failure.

MATERIALS AND METHODS: Imaging records were searched for patients ≤18 years of age who underwent 2-D gradient recalled echo (GRE) MR elastography of the liver between September 2011 and August 2015 on one of two 1.5-T MRI platforms. Imaging reports and clinical records were reviewed for failed MR elastography acquisitions, factor(s) resulting in failure and whether a subsequent successful examination had been performed.

RESULTS: Four hundred sixty-eight MR elastography examinations were performed in 372 patients between 1.5 months and 18 years of age during the study period. Ninety-six percent (450/468) of the examinations were successful. There was no significant difference in mean age (12.6±3.6 vs. 11.2±4.1 years, P=0.12) or body mass index (BMI) (28.2±12.4 vs. 29.5±10 kg/m2, P=0.6) between patients with and without successful examinations. MR elastography failures were due to poor paddle positioning resulting in inadequate generation of hepatic shear waves (n=5), iron overload (n=4), patient inability to tolerate MRI (n=3), patient breathing/motion (n=3), artifact from implanted hardware (n=1) and technical malfunction (n=2). Seven of nine (78%) repeat examinations were successful (78%).

CONCLUSION: Hepatic 2-D GRE MR elastography at 1.5 T is technically robust in children. Exam failure is infrequent and largely reflects patient specific factors, some of which can be mitigated with careful technique.

来源:Pediatr Radiol,查看原文。版权归原作者所有。

引用

  1. Joshi M, Dillman JR, Towbin AJ, et al. MR elastography: high rate of technical success in pediatric and young adult patients. Pediatr Radiol. 2017. PMID: 28367603. doi:10.1007/s00247-017-3831-z

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。