利用超声衰减成像监测离体牛肝微波消融
本研究对11例离体牛肝微波消融操作生成超声衰减系数图,消融区平均衰减系数为0.9 (±0.2) dB/cm/MHz,未消融区为0.45 (±0.15) dB/cm/MHz,衰减图像与病理图像相关性良好。结果表明无创衰减成像可能有助于监测肝脏热消融治疗。
要点
- 消融区平均衰减系数0.9 (±0.2) dB/cm/MHz,高于未消融区0.45 (±0.15) dB/cm/MHz。
- 超声衰减图像与消融区病理图像相关性良好。
- B型超声对消融区边界显示效果有限,衰减成像可提供定量补充信息。
核心结论
- 超声衰减成像生成的衰减系数图与离体牛肝微波消融区的病理图像相关性良好。
- 消融区平均衰减系数为0.9 (±0.2) dB/cm/MHz,未消融区为0.45 (±0.15) dB/cm/MHz。
- 上述估计值与文献报道值一致,提示无创衰减成像对监测肝脏热消融治疗可能有用。
研究设计与样本
- 研究类型为离体牛肝实验,对11例微波消融操作生成衰减系数图。
- 使用先前引入的衰减估计方法生成定量衰减系数图。
- 以病理图像作为参考标准进行相关性比较。
关键影像表现/征象
- B型超声因消融区与周围正常组织回声对比度差,难以清晰勾画消融区边界。
- 定量超声可提供背向散射系数和衰减系数等声学特性估计,更有效地区分消融与未消融区域。
- 衰减图像与消融区病理图像相关性良好。
诊断标准与关键数值
- 消融区感兴趣区平均衰减系数为0.9 (±0.2) dB/cm/MHz。
- 未消融区感兴趣区平均衰减系数为0.45 (±0.15) dB/cm/MHz。
临床应用与阅片要点
- 衰减系数图可作为B型超声的补充,用于区分消融区与未消融区。
- 消融区衰减系数高于未消融区,阅片时可关注衰减图上对应区域的高值区。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Thermal ablation of soft tissue changes the tissue microstructure and, consequently, induces changes in its acoustic properties. Although B-mode ultrasound provides high-resolution and high-frame-rate images of ablative therapeutic procedures, it is not particularly effective at delineating boundaries of ablated regions because of poor contrast in echogenicity between ablated and surrounding normal tissue. Quantitative ultrasound techniques can provide quantitative estimates of acoustic properties, such as backscatter and attenuation coefficients, and differentiate ablated and unablated regions more effectively, with the potential for monitoring minimally invasive thermal therapies. In this study, a previously introduced attenuation estimation method was used to create quantitative attenuation coefficient maps for 11 microwave ablation procedures performed on refrigerated ex vivo bovine liver. The attenuation images correlate well with the pathologic images of the ablated region. The mean attenuation coefficient for regions of interest drawn inside and outside the ablated zones were 0.9 (±0.2) and 0.45 (±0.15) dB/cm/MHz, respectively. These estimates agree with reported values in the literature and establish the usefulness of non-invasive attenuation imaging for monitoring therapeutic procedures in the liver.