对比增强CT冠状动脉造影中采用厂商特定模型迭代重建算法改进冠状动脉斑块与管腔衰减的评估
研究利用冠状动脉血管体模和10例临床患者,比较滤波反投影与两种迭代重建算法对斑块衰减稳定性和冠状动脉管腔可视化的影响。结果显示FIRST算法使斑块CT值随对比增强的变化更小,管腔视觉评分更高,有助于改善冠状动脉斑块显示。
要点
- FIRST重建下斑块CT值随对比增强的变化为-12至8 HU,小于FBP的15-44 HU和AIDR3D的10-31 HU。
- 血管腔视觉评分FIRST为3.7±0.5,高于FBP的2.8±0.6和AIDR3D的3.5±0.5。
- FIRST算法控制斑块密度增加和管腔-斑块对比,改善冠状动脉CT造影中斑块的可视化。
核心结论
- FIRST方法控制斑块密度增加和管腔-斑块对比,从而改善冠状动脉CT血管造影中冠状动脉斑块的可视化。
研究设计与样本
- 研究采用冠状动脉血管体模和10例临床患者,使用320排容积扫描,比较FBP、AIDR3D和FIRST三种重建算法对斑块衰减和管腔可视化的影响,参考标准为体模设定及4分法视觉评分。
关键影像表现/征象
- 在50%狭窄时,FBP和AIDR3D的斑块CT值随对比增强而增加,FIRST的斑块CT值变化较小。
- 血管腔视觉评分在FBP、AIDR3D和FIRST分别为2.8±0.6、3.5±0.5和3.7±0.5。
诊断标准与关键数值
- 体模狭窄程度为50%和75%,斑块CT值约40、80、150 HU,管腔目标CT值约350 HU、450 HU和0 HU。
- 有和无对比增强时斑块CT值差值:FBP为15-44 HU,AIDR3D为10-31 HU,FIRST为-12至8 HU。
- 临床图像质量采用4分法评估冠状动脉管腔和斑块。
临床应用与阅片要点
- FIRST重建可减少对比增强对斑块CT值的影响,有助于稳定评估斑块衰减。
- FIRST重建下血管腔视觉评分较高,可能改善冠状动脉管腔的显示。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: To investigate the stabilities of plaque attenuation and coronary lumen for different plaque types, stenotic degrees, lumen densities, and reconstruction methods using coronary vessel phantoms and the visualization of coronary plaques in clinical patients through coronary computed tomography (CT) angiography.
MATERIALS AND METHODS: We performed 320-detector volume scanning of vessel tubes with stenosis and a tube without stenosis using three types of plaque CT numbers. The stenotic degrees were 50% and 75%. Images were reconstructed with filtered back projection (FBP) and two types of iterative reconstructions (AIDR3D and FIRST [forward-projected model-based iterative reconstruction solution]), with stenotic CT number of approximately 40, 80, and 150 HU (Hounsfield unit), respectively. In each case, the tubing of the coronary vessel was filled with diluted contrast material and distilled water to reach the target lumen CT numbers of approximately 350 HU and 450 HU, and 0 HU, respectively. Peak lumen and plaque CT numbers were measured to calculate the lumen-plaque contrast. In addition, we retrospectively evaluated the image quality with regard to coronary arterial lumen and the plaque in 10 clinical patients on a 4-point scale.
RESULTS: At 50% stenosis, the plaque CT number with contrast enhancement increased for FBP and AIDR3D, and the difference in the plaque CT number with and without contrast enhancement was 15-44 HU for FBP and 10-31 HU for AIDR3D. However, the plaque CT number for FIRST had a smaller variation and the difference with and without contrast enhancement was -12 to 8 HU. The visual evaluation score for the vessel lumen was 2.8 ± 0.6, 3.5 ± 0.5, and 3.7 ± 0.5 for FBP, AIDR3D, and FIRST, respectively.
CONCLUSIONS: The FIRST method controls the increase in plaque density and the lumen-plaque contrast. Consequently, it improves the visualization of coronary plaques in coronary CT angiography.
引用
- Funama Y, Utsunomiya D, Hirata K, et al. Improved Estimation of Coronary Plaque and Luminal Attenuation Using a Vendor-specific Model-based Iterative Reconstruction Algorithm in Contrast-enhanced CT Coronary Angiography. Acad Radiol. 2017. PMID: 28396126. doi:10.1016/j.acra.2017.02.006
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。