超低剂量CT评估冠状动脉钙化的性能:与冠状动脉钙化评分的直接比较
研究对74例因评估冠状动脉疾病转诊行冠状动脉CT血管成像的患者同期行标准冠状动脉钙化评分采集与超低剂量CT,超低剂量CT采用四点法主观分级并计算总钙化评分。超低剂量CT检出冠状动脉钙化的敏感度80.9%、特异度85.2%、准确率82.4%,AUC为0.87,辐射剂量显著低于标准采集;对于至少中度钙负荷(Agatston评分>100)者,超低剂量CT不太可能漏诊冠状动脉钙化。
要点
- 超低剂量CT总钙化评分≥1检出冠状动脉钙化的敏感度80.9%、特异度85.2%、准确率82.4%、AUC 0.87。
- Agatston评分>100的至少中度钙负荷者,超低剂量CT不太可能漏诊冠状动脉钙化。
- 超低剂量CT的平均DLP为9.3 mGy·cm,显著低于标准钙化评分采集的77.7 mGy·cm。
核心结论
- 超低剂量CT对冠状动脉钙化检出具有良好敏感度、特异度和总体准确率,且辐射剂量显著低于标准冠状动脉钙化评分采集;对于至少中度钙负荷(Agatston评分>100)的个体,超低剂量CT不太可能漏诊冠状动脉钙化。
研究设计与样本
- 研究对因评估冠状动脉疾病转诊行冠状动脉CT血管成像的患者同期行标准冠状动脉钙化评分采集和超低剂量CT,共74例。
诊断标准与关键数值
- 超低剂量CT对全部研究及各冠状动脉节段钙化采用四点法主观分级(0为无钙化,3为严重钙化),各节段评分之和为超低剂量CT总钙化评分。超低剂量CT总钙化评分≥1检出冠状动脉钙化的敏感度80.9%、特异度85.2%、准确率82.4%,受试者工作特征曲线下面积0.87;标准冠状动脉钙化评分采集检出钙化47例(63.5%),中位Agatston评分41(四分位距0–263)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
AIM: To evaluate the diagnostic performance of ultra-low-dose computed tomography (ULDCT) in comparison to standard coronary calcium score (CCS) acquisition for the evaluation of coronary artery calcification (CAC).
MATERIALS AND METHODS: Standard CCS acquisition and ULDCT were performed in patients referred for coronary CT angiography for the evaluation of coronary artery disease. CAC in ULDCT was graded subjectively using a four-point scale (from 0, no calcification, to 3, severe calcification) for the complete study and for each individual coronary segment. The summation of all individual coronary segment scores generated an ULDCT total CAC score. ULDCT results were compared to standard Agatston score and sensitivity and specificity of ULDCT were calculated.
RESULTS: CCS and ULDCT were performed in 74 patients, with a mean DLP of 77.7 mGy·cm (±12.1) and 9.3 mGy·cm (±0.6), respectively (p<0.001). Coronary calcification was detected in 47 patients (63.5%) in standard CCS acquisition (median Agatston score of 41; interquartile range [IQR]:0263), in comparison to 42 patients (56.8%) in ULDCT (p<0.001). The sensitivity and specificity of the ULDCT total CAC score ≥1 was 80.9% and 85.2%, respectively, with an accuracy of 82.4%. The area under the receiver operating characteristic curve for the presence of CAC was 0.87.
CONCLUSION: ULDCT shows good sensitivity, specificity, and overall accuracy for the detection of coronary calcification with a markedly lower radiation dose in comparison to CCS. ULDCT is unlikely to miss coronary calcification in individuals with at least moderate calcium load (Agatston score >100).
引用
- Silveira de Souza VV, Soares Torres F, Hochhegger B, et al. Performance of ultra-low-dose CT for the evaluation of coronary calcification: a direct comparison with coronary calcium score. Clin Radiol. 2017. PMID: 28413071. doi:10.1016/j.crad.2017.03.015