左心室辅助装置转速测试中左心室与右心室的三维形态学变化
本研究对31例左心室辅助装置(LVAD)患者(19例HeartMate II、12例HVAD)在转速测试中同步行三维超声心动图和右心导管检查,分析左、右心室容积与形态变化。结果显示两种装置随转速增加心输出量升高、楔压下降,但左、右心室容积和形态反应不同,HMII变化更明显。作者认为在临床超声转速测试中加入右心室评估可能更好地优化LVAD转速,尚需进一步研究其对患者结局的影响。
要点
- LVAD转速测试中应同时评估左、右心室容积与形态,而非仅看左心室。
- HMII随转速增加左心室容积平均减少127 ml、右心室容积平均增加60 ml。
- HVAD变化相似但较轻,左心室在高转速下仍显著更接近球形。
核心结论
- 两种LVAD对转速变化的心脏反应不同,体现在左、右心室的容积和形态改变上;在临床超声转速测试中加入右心室评估可能更好地优化LVAD转速,尚需进一步研究确定其对患者结局的影响。
研究设计与样本
- 研究论文;31例LVAD患者(19例HMII、12例HVAD),同步行三维超声心动图、转速测试和右心导管检查,转速逐步增加(HMII 8,000至12,000转/分,HVAD 2,300至3,200转/分),采集三维全容积左、右心室数据集,用定制软件分析心内膜面计算左心室球形度、锥度(完美球/锥=1)及右心室室间隔和游离壁曲率(0=平坦;<0=凹;>0=凸)。
关键影像表现/征象
- 两种装置随转速增加心输出量升高、楔压下降。
- HMII左心室容积进行性减少(平均Δ=127 ml),左心室变得不那么球形、更呈锥形;右心室容积初期稳定,随后在较高转速下增加(平均Δ=60 ml)。
- HVAD表现相似但较轻(左心室平均Δ=51 ml,右心室平均Δ=22 ml),且左心室即使在高转速下仍显著更呈球形。
- HMII患者在最高转速时右心室室间隔变得更凸(凸向左心室),而HVAD患者右心室室间隔无明显变化。
诊断标准与关键数值
- 左心室球形度和锥度以完美球/锥=1为参照。
- 右心室室间隔和游离壁曲率以0=平坦、<0=凹、>0=凸为判定标准。
临床应用与阅片要点
- 临床超声转速测试中可考虑在左心室评估之外加入右心室容积与形态评估,以更好地优化LVAD转速(作者建议)。
- HMII与HVAD的转速范围和形态反应不同,解读时应结合装置类型。
局限性
- 原文结论指出尚需进一步研究确定加入右心室评估是否对患者结局有影响。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: The purpose of this study was to investigate the differential impact of the 2 most commonly available left ventricular assist device (LVAD) types on the right ventricle (RV) and left ventricle (LV) using 3-dimensional (3D) echocardiography-based analysis of ventricular morphology.
BACKGROUND: LVADs have emerged as common therapy for advanced heart failure. Recent data suggest that the heart responds differently to speed settings in the 2 main devices available (HeartMate II [HMII], St Jude Medical, Pleasanton, California, and HVAD, HeartWare International, Framingham, Massachusetts). The authors hypothesized that 3D echocardiographic assessment of LV and RV volumes and shape would help describe the differential impact of the 2 LVAD types on the heart.
METHODS: Simultaneous 3D echocardiography, ramp test, and right heart catheterization were performed in 31 patients with LVADs (19 with HMII and 12 with HVAD). Device speed was increased stepwise (8,000 to 12,000 for HMII and 2,300 to 3,200 revolutions per minute for HVAD). 3D echocardiographic full-volume LV and RV datasets were acquired, and endocardial surfaces were analyzed using custom software to calculate LV sphericity, conicity (perfect sphere/cone = 1) and RV septal and free-wall curvature (0 = flat; <0 = concave; >0 = convex).
RESULTS: For both devices, cardiac output increased and wedge pressure decreased with increasing speed. In HMII, LV volumes progressively decreased (meanΔ = 127 ml) as the LV became less spherical and more conical, whereas the RV volume initially remained stable, but subsequently increased at higher speeds (meanΔ = 60 ml). Findings for the HVAD were similar, but less pronounced (LV:meanΔ = 51 ml, RV:meanΔ = 22 ml), and the LV remained significantly more spherical even at high speeds. On average, in HMII patients, the RV septum became more convex (bulging into the LV) at the highest speeds whereas in HVAD patients, there was no discernable change in the RV septum.
CONCLUSIONS: The heart responds differently to pump speed changes with the 2 types of LVAD, as reflected by the volume and shape changes of both the LV and RV. Our study suggests that adding RV assessment to the clinical echo-ramp study may better optimize LVAD speed. Further study is needed to determine whether this would have an impact on patient outcomes.