动力注射与手动注射对对比增强磁共振血管成像团注形态及图像质量的比较:猪模型实验研究
在12头猪中比较动力注射与手动注射钆布醇(0.1 mmol/kg,2 mL/s)对CE-MRA团注形态和图像质量的影响。动力注射的3D MRA信噪比更高(P < 0.001),12次扫描均呈标准团注曲线而手动注射仅1次,图像质量(P = 0.001)及4D MRA团注过渡可见性(P = 0.031)均更优。
要点
- 动力注射3D MRA信噪比显著高于手动注射(P < 0.001)。
- 动力注射12次扫描均呈标准团注曲线,手动注射仅1次。
- 动力注射图像质量更高(P = 0.001),动脉强化更高(P = 0.031)且非目标强化重叠更少(P = 0.001)。
核心结论
- 与手动注射相比,动力注射提供更标准化的团注形态和更高的图像质量,源于更高的动脉强化和更少的非目标血管重叠。
研究设计与样本
- 实验研究,12头猪,每头猪接受动力注射和手动注射钆布醇(0.1 mmol/kg,2 mL/s),每种CE-MRA类型各6组动力与手动注射。
关键影像表现/征象
- 动力注射在3D MRA上信噪比显著高于手动注射(P < 0.001)。
- 动力注射12次扫描均呈标准团注曲线(尖锐首过峰后接平缓再循环峰),手动注射仅1/12次。
- 动力注射达峰时间标准差小于手动注射。
- 动力注射动脉强化显著更高(P = 0.031),非目标强化重叠显著更少(P = 0.001)。
- 动力注射4D MRA团注过渡可见性显著优于手动注射(P = 0.031)。
临床应用与阅片要点
- 两种注射方法在3D MRA上均达到诊断质量,但动力注射图像质量显著更高(P = 0.001)。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The aim of this study was to compare power versus manual injection in bolus shape and image quality on contrast-enhanced magnetic resonance angiography (CE-MRA).
MATERIALS AND METHODS: Three types of CE-MRA (head-neck 3-dimensional [3D] MRA with a test-bolus technique, thoracic-abdominal 3D MRA with a bolus-tracking technique, and thoracic-abdominal time-resolved 4-dimensional [4D] MRA) were performed after power and manual injection of gadobutrol (0.1 mmol/kg) at 2 mL/s in 12 pigs (6 sets of power and manual injections for each type of CE-MRA). For the quantitative analysis, the signal-to-noise ratio was measured on ascending aorta, descending aorta, brachiocephalic trunk, common carotid artery, and external carotid artery on the 6 sets of head-neck 3D MRA, and on ascending aorta, descending aorta, brachiocephalic trunk, abdominal aorta, celiac trunk, and renal artery on the 6 sets of thoracic-abdominal 3D MRA. Bolus shapes were evaluated on the 6 sets each of test-bolus scans and 4D MRA. For the qualitative analysis, arterial enhancement, superimposition of nontargeted enhancement, and overall image quality were evaluated on 3D MRA. Visibility of bolus transition was assessed on 4D MRA. Intraindividual comparison between power and manual injection was made by paired t test, Wilcoxon rank sum test, and analysis of variance by ranks.
RESULTS: Signal-to-noise ratio on 3D MRA was statistically higher with power injection than with manual injection (P < 0.001). Bolus shapes (test-bolus, 4D MRA) were represented by a characteristic standard bolus curve (sharp first-pass peak followed by a gentle recirculation peak) in all the 12 scans with power injection, but only in 1 of the 12 scans with manual injection. Standard deviations of time-to-peak enhancement were smaller in power injection than in manual injection. Qualitatively, although both injection methods achieved diagnostic quality on 3D MRA, power injection exhibited significantly higher image quality than manual injection (P = 0.001) due to significantly higher arterial enhancement (P = 0.031) and less superimposition of nontargeted enhancement (P = 0.001). Visibility of bolus transition on 4D MRA was significantly better with power injection than with manual injection (P = 0.031).
CONCLUSIONS: Compared with manual injection, power injection provides more standardized bolus shapes and higher image quality due to higher arterial enhancement and less superimposition of nontargeted vessels.