用于放疗治疗计划的CT扫描协议能否调整以优化图像质量和患者剂量?一项系统综述
本系统综述检索PubMed和EMBASE并追溯参考文献,纳入53篇相关文献,评估CT扫描协议调整对CT值及放疗计划剂量的影响。CT扫描参数显著影响图像质量,管电压及部分扫描仪的视野和重建算法可致CT值明显变化。软组织CT值变化对计划剂量影响大于骨和空气,建议使用体模评估CT值变化,并提示软组织±20 HU、肺和骨±50 HU的容差可将计划剂量变化限制在<1%。
要点
- CT扫描参数显著影响图像质量,管电压及部分扫描仪的视野和重建算法可致CT值明显变化。
- 软组织CT值变化对计划剂量的影响大于骨和空气CT值变化。
- 评估CT值变化时建议使用人体模体;软组织±20 HU、肺和骨±50 HU的容差可能将计划剂量变化限制在<1%。
核心结论
- CT扫描协议调整会影响CT图像质量和CT值,部分参数变化可致CT值明显改变;软组织CT值变化对放疗计划剂量的影响大于骨和空气,文献提示软组织±20 HU、肺和骨±50 HU的容差可将计划剂量变化限制在<1%,但放疗CT扫描协议优化的已发表研究有限。
研究设计与样本
- 系统综述,检索PubMed和EMBASE并追溯参考文献,共纳入53篇相关文献。
诊断标准与关键数值
- 文献提示软组织CT值容差为±20 HU、肺和骨CT值容差为±50 HU时,可将治疗计划剂量变化限制在<1%。
临床应用与阅片要点
- 评估CT值变化时建议使用人体模体。
- CT值变化可耐受程度取决于身体部位和大小、计划算法、治疗束能量及计划类型。
局限性
- 关于放疗中CT扫描协议优化的已发表研究有限。
- 扫描参数对HU值的影响未被全面报道。
- 结论已呈现并明确了进一步研究方向。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
This article reviews publications related to the use of CT scans for radiotherapy treatment planning, specifically the impact of scan protocol changes on CT number and treatment planning dosimetry and on CT image quality. A search on PubMed and EMBASE and a subsequent review of references yielded 53 relevant articles. CT scan parameters significantly affect image quality. Some will also affect Hounsfield unit (HU) values, though this is not comprehensively reported on. Changes in tube kilovoltage and, on some scanners, field of view and reconstruction algorithms have been found to produce notable HU changes. The degree of HU change which can be tolerated without changing planning dose by >1% depends on the body region and size, planning algorithms, treatment beam energy and type of plan. A change in soft-tissue HU value has a greater impact than changes in HU for bone and air. The use of anthropomorphic phantoms is recommended when assessing HU changes. There is limited published work on CT scan protocol optimization in radiotherapy. Publications suggest that HU tolerances of ±20 HU for soft tissue and of ±50 HU for the lung and bone would restrict dose changes in the treatment plan to <1%. Literature related to the use of CT images in radiotherapy planning has been reviewed to establish the acceptable level of HU change and the impact on image quality of scan protocol adjustment. Conclusions have been presented and further work identified.
引用
- Davis AT, Palmer AL, Nisbet A, et al. Can CT scan protocols used for radiotherapy treatment planning be adjusted to optimize image quality and patient dose? A systematic review. Br J Radiol. 2017. PMID: 28452568. doi:10.1259/bjr.20160406
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。