利用三维旋转透视确定方向性脑深部刺激电极的方位
研究将两根方向性电极插入3D打印石膏颅骨(填充明胶),用三维旋转透视评估电极方位;3名盲法评估者分析12例随机方位透视图像,显示考虑X射线标记时一致性界限为±9.37°、组内相关系数0.9975,并考虑电极段间隙重叠线时一致性界限为±2.44°、组内相关系数0.9998。结论认为三维旋转透视联合所述评估方法可确定电极精确方位。
要点
- 方向性电极的精确方位可通过三维旋转透视结合X射线标记与电极段间隙重叠线来判定。
- 仅用X射线标记时,评估者间一致性界限为±9.37°、组内相关系数0.9975。
- 结合电极段间隙重叠线后,一致性界限缩小至±2.44°、组内相关系数0.9998。
核心结论
- 在方向性脑深部刺激系统中,三维旋转透视联合所述评估方法可确定电极的精确方位,从而发挥影像辅助个体化编程的全部潜力。
研究设计与样本
- 体外研究:将两根方向性电极插入填充明胶的3D打印石膏颅骨,先研究电极尖端相对于钻孔层面电极的扭转,再由3名盲法评估者评估12例随机电极方位的三维旋转透视图像。
关键影像表现/征象
- 可显示电极扭转达35°。
- 仅考虑X射线标记评估电极旋转时,一致性界限为±9.37°,组内相关系数为0.9975。
- 同时考虑电极段间隙重叠形成的线时,与均值的一致性界限为±2.44°,组内相关系数为0.9998。
诊断标准与关键数值
- 仅考虑X射线标记时,评估者间一致性界限±9.37°,组内相关系数0.9975。
- 考虑电极段间隙重叠线时,评估者间一致性界限±2.44°,组内相关系数0.9998。
临床应用与阅片要点
- 方向性刺激的影像辅助编程需先明确电极阵列的精确方位。
- 三维旋转透视评估电极方位时,除X射线标记外可结合电极段间隙重叠线提高评估者间一致性。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: New deep brain stimulation leads with electrode contacts that are split along their circumference allow steering of the electrical field in a predefined direction. However, imaging-assisted directional stimulation requires detailed knowledge of the exact orientation of the electrode array. The purpose of this study was to evaluate whether this information can be obtained by rotational 3D fluoroscopy.
MATERIALS AND METHODS: Two directional leads were inserted into a 3D-printed plaster skull filled with gelatin. The torsion of the lead tip versus the lead at the burr-hole level was investigated. Then, 3 blinded raters evaluated 12 3D fluoroscopies with random lead orientations. They determined the lead orientation considering the x-ray marker only and considering the overlap of the gaps between the contact segments. Intraclass correlation coefficients and an extended version of the Bland-Altman plot were used to determine interrater reliability and agreement of the measurements of the different raters.
RESULTS: Electrode torsion of up to 35° could be demonstrated. Evaluation of the lead rotation considering the x-ray marker only revealed limits of agreement of ±9.37° and an intraclass correlation coefficient of 0.9975. In addition, taking into account the lines resulting from overlapping of the gaps between the electrode segments, the limits of agreement to the mean were ±2.44° and an intraclass correlation coefficient of 0.9998.
CONCLUSIONS: In directional deep brain stimulation systems, rotational 3D fluoroscopy combined with the described evaluation method allows for determining the exact orientation of the leads, enabling the full potential of imaging-assisted personalized programming.
引用
- Reinacher PC, Krüger MT, Coenen VA, et al. Determining the Orientation of Directional Deep Brain Stimulation Electrodes Using 3D Rotational Fluoroscopy. AJNR Am J Neuroradiol. 2017. PMID: 28385887. doi:10.3174/ajnr.A5153
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。