文献速递神经外周血管MRDSA

零回波时间MR血管成像与飞行时间MR血管成像用于弹簧圈栓塞颅内动脉瘤随访评估的初步图像质量研究

Follow-up assessment of coiled intracranial aneurysms using zTE MRA as compared with TOF MRA: a preliminary image quality study.
PubMed 文献 · Eur Radiol · 2017-04-05

前瞻性纳入25例弹簧圈栓塞颅内动脉瘤患者,在DSA前24小时行零回波时间(zTE)MRA和飞行时间(TOF)MRA,以DSA为参考评估图像质量和闭塞状态。zTE在磁化率伪影强度和载瘤动脉血流信号方面评分均高于TOF,与DSA相关性更高(Spearman ρ 0.91 vs 0.81),zTE MRA在随访评估中优于TOF MRA。

要点

  • zTE MRA对磁化率伪影和血流动力学的敏感性低于TOF MRA,载瘤动脉显示更优。
  • zTE MRA评估闭塞状态与DSA一致性极好(加权κ=0.95),优于TOF MRA(加权κ=0.80)。
  • zTE MRA与DSA相关性高(Spearman ρ=0.91),有望成为TOF MRA的替代方案。

核心结论

  • zTE MRA在弹簧圈栓塞颅内动脉瘤随访评估中显示出良好前景,在显示载瘤动脉和评估闭塞状态方面优于TOF MRA,在本单中心研究中与DSA等效。

研究设计与样本

  • 前瞻性单中心研究,纳入25例患者(男10例、女15例,年龄53.96±12.46岁),在DSA前24小时行zTE MRA和TOF MRA,以DSA为参考标准。

关键影像表现/征象

  • zTE的磁化率伪影强度评分高于TOF(3.42±0.64 vs 2.92±0.63,P=0.01)。
  • zTE的载瘤动脉血流信号评分高于TOF(3.66±0.95 vs 3.24±1.24,P=0.01)。
  • DSA显示完全闭塞17例、残留瘤颈5例、残留动脉瘤2例。

诊断标准与关键数值

  • 磁化率伪影强度和载瘤动脉血流信号均采用4分制评分。
  • 闭塞状态采用3级Montreal量表评估。
  • zTE观察者间一致性极好(加权κ=0.89),TOF观察者间一致性良好(加权κ=0.68)。
  • zTE与DSA的检查方法间一致性极好(加权κ=0.95),TOF与DSA的检查方法间一致性良好(加权κ=0.80)。
  • zTE与DSA相关性高(Spearman ρ=0.91),TOF与DSA相关性高(Spearman ρ=0.81)。

临床应用与阅片要点

  • zTE MRA对磁化率伪影和血流动力学较不敏感,有助于显示弹簧圈栓塞后的载瘤动脉。
  • zTE MRA评估闭塞状态与DSA一致性极好,可作为TOF MRA的替代方案用于随访评估。

局限性

  • 原文标题注明为初步图像质量研究。
  • 原文结论称zTE MRA显示出良好前景,zTE MRA可能为TOF MRA的替代方案。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To prospectively assess coiled intracranial aneurysms using a novel non-contrast enhanced zero echo time (zTE) MR angiography (MRA) method, and compare its image quality with time-of-flight (TOF) MRA, using digital subtraction angiography (DSA) as reference.

METHODS: Twenty-five patients (10 males and 15 females; age 53.96 ± 12.46 years) were enrolled in this monocentric study. MRA sequences were performed 24 h before DSA. Susceptibility artefact intensity and flow signal within the parent artery were carried out using a 4-point scale. Occlusion status was assessed using the 3-grade Montreal scale.

RESULTS: Scores of zTE were higher than TOF for both susceptibility artefact intensity (3.42 ± 0.64, 2.92 ± 0.63, P = 0.01) and flow signal (3.66 ± 0.95, 3.24 ± 1.24, P = 0.01). DSA revealed 17 complete occlusions, five residual neck aneurysms and two residual aneurysms. Inter-observer agreement was excellent (weighted κ: 0.89) for zTE and good (weighted κ: 0.68) for TOF. Intermodality agreement was excellent for zTE (weighted κ: 0.95) and good for TOF (weighted κ: 0.80). Correlations of both MRA sequences with DSA were high (zTE, Spearman's ρ: 0.91; TOF, Spearman's ρ: 0.81).

CONCLUSIONS: zTE MRA showed promising results for follow-up assessment of coiled intracranial aneurysms and was superior to TOF MRA for visualizing the parent artery and evaluating occlusion status.

KEY POINTS: • Various MRA sequences were applied for follow-up assessment of coiled intracranial aneurysms. • zTE MRA was less sensitive to susceptibility artefacts and haemodynamics. • In this monocentric study, zTE MRA was equivalent to DSA. • zTE MRA maybe an alternative to TOF MRA for follow-up assessment.

来源:Eur Radiol,查看原文。版权归原作者所有。