文献速递神经MR

合成MRI用于临床神经影像:磁共振图像合成(MAGiC)前瞻性、多中心、多阅片者试验结果

Synthetic MRI for Clinical Neuroimaging: Results of the Magnetic Resonance Image Compilation (MAGiC) Prospective, Multicenter, Multireader Trial.
PubMed 文献 · AJNR Am J Neuroradiol · 2017-04-27

该前瞻性多阅片者多病例非劣效性试验纳入109例受试者的1526幅图像,由7名盲法神经放射科医生评估,比较合成与常规脑MRI的诊断质量、形态可读性、影像学发现及伪影。合成MRI整体诊断质量非劣效于常规MRI(P<.001),合成T2 FLAIR伪影更常见但可识别且不模仿病变。

要点

  • 合成MRI整体诊断质量非劣效于常规MRI(P<.001),平均合成-常规差值为-0.335±0.352。
  • 合成与常规形态可读性一致率>95%,仅内囊后肢在T1、T1 FLAIR和质子密度视图一致率>80%。
  • 合成T2 FLAIR伪影更明显:血流伪影增加24.1%,白噪声伪影增加17.6%的病例。

核心结论

  • 合成MRI整体质量与常规质子密度、STIR及T1、T2加权对比视图在神经系统疾病中相似。
  • 合成T2 FLAIR伪影更常见,但易于识别且不模仿病变,可能需额外常规T2 FLAIR确认诊断。

研究设计与样本

  • 前瞻性多阅片者、多病例非劣效性试验,纳入109例受试者(平均53.0±18.5岁,范围19-89岁)的1526幅图像。
  • 由7名盲法神经放射科医生独立评估,图像随机化。
  • 每个病例包括常规T1加权、T2加权、T1 FLAIR、T2 FLAIR、STIR和/或质子密度及来自多动态多回波成像的合成重建图像。

关键影像表现/征象

  • 临床MR检查显示46例健康病例和63例病理病例。
  • 合成T2 FLAIR有更明显的伪影,包括血流伪影增加24.1%的病例和白噪声伪影增加17.6%的病例。

诊断标准与关键数值

  • 整体诊断质量在5级Likert量表上合成MRI非劣效于常规成像(P<.001;平均合成-常规差值为-0.335±0.352;Δ=0.5;95%CI下限为-0.402)。
  • 合成与常规形态可读性一致率>95%,但内囊后肢在T1、T1 FLAIR和质子密度视图一致率>80%。

临床应用与阅片要点

  • 合成MRI整体诊断质量与常规质子密度、STIR及T1、T2加权对比视图相似。
  • 合成T2 FLAIR伪影更常见,但易于识别且不模仿病变,可能需额外常规T2 FLAIR确认诊断。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND AND PURPOSE: Synthetic MR imaging enables reconstruction of various image contrasts from 1 scan, reducing scan times and potentially providing novel information. This study is the first large, prospective comparison of synthetic-versus-conventional MR imaging for routine neuroimaging.

MATERIALS AND METHODS: A prospective multireader, multicase noninferiority trial of 1526 images read by 7 blinded neuroradiologists was performed with prospectively acquired synthetic and conventional brain MR imaging case-control pairs from 109 subjects (mean, 53.0 ± 18.5 years of age; range, 19-89 years of age) with neuroimaging indications. Each case included conventional T1- and T2-weighted, T1 and T2 FLAIR, and STIR and/or proton density and synthetic reconstructions from multiple-dynamic multiple-echo imaging. Images were randomized and independently assessed for diagnostic quality, morphologic legibility, radiologic findings indicative of diagnosis, and artifacts.

RESULTS: Clinical MR imaging studies revealed 46 healthy and 63 pathologic cases. Overall diagnostic quality of synthetic MR images was noninferior to conventional imaging on a 5-level Likert scale (P < .001; mean synthetic-conventional, -0.335 ± 0.352; Δ = 0.5; lower limit of the 95% CI, -0.402). Legibility of synthetic and conventional morphology agreed in >95%, except in the posterior limb of the internal capsule for T1, T1 FLAIR, and proton-density views (all, >80%). Synthetic T2 FLAIR had more pronounced artifacts, including +24.1% of cases with flow artifacts and +17.6% cases with white noise artifacts.

CONCLUSIONS: Overall synthetic MR imaging quality was similar to that of conventional proton-density, STIR, and T1- and T2-weighted contrast views across neurologic conditions. While artifacts were more common in synthetic T2 FLAIR, these were readily recognizable and did not mimic pathology but could necessitate additional conventional T2 FLAIR to confirm the diagnosis.

来源:AJNR Am J Neuroradiol,查看原文。版权归原作者所有。

引用

  1. Tanenbaum LN, Tsiouris AJ, Johnson AN, et al. Synthetic MRI for Clinical Neuroimaging: Results of the Magnetic Resonance Image Compilation (MAGiC) Prospective, Multicenter, Multireader Trial. AJNR Am J Neuroradiol. 2017. PMID: 28450439. doi:10.3174/ajnr.A5227

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。