文献速递神经技术与安全MR

利用临床影像档案进行影像组学研究:脑体积自动测量方法的可靠性

Leveraging Clinical Imaging Archives for Radiomics: Reliability of Automated Methods for Brain Volume Measurement.
PubMed 文献 · Radiology · 2017-04-27

本研究利用临床影像档案中同一38例患者(年龄1-71岁,平均22岁,女性11例)的薄层科研级与常规厚层临床脑MR图像,验证SPM、FreeSurfer和FSL三种自动工具包估算全脑体积、灰质体积和白质体积的可靠性。SPM在厚层图像上估算全脑体积、灰质体积和白质体积的组内相关系数分别为0.97、0.85和0.83,表现优异;FSL相应组内相关系数分别为0.69、0.51和0.60,低于SPM;FreeSurfer仅全脑体积组内相关系数为0.63。结论认为厚层临床质量MR图像可借助SPM可靠用于计算全脑体积、灰质体积和白质体积等定量脑指标。

要点

  • SPM在厚层临床MR图像上估算全脑体积的组内相关系数为0.97,灰质体积0.85,白质体积0.83。
  • FSL估算全脑体积、灰质体积和白质体积的组内相关系数分别为0.69、0.51和0.60,低于SPM。
  • FreeSurfer仅全脑体积组内相关系数为0.63,灰质体积和白质体积未报告。

核心结论

  • 厚层临床质量MR图像可借助SPM可靠用于计算全脑体积、灰质体积和白质体积等定量脑指标。

研究设计与样本

  • 方法学验证研究,使用临床影像档案中同一38例患者(年龄1-71岁,平均22岁,女性11例)的薄层科研级与常规厚层临床脑MR图像,应用SPM、FreeSurfer和FSL三种自动工具包估算脑体积,以薄层图像为参照计算组内相关系数。

诊断标准与关键数值

  • SPM估算全脑体积、灰质体积和白质体积的组内相关系数分别为0.97、0.85和0.83。
  • FSL估算全脑体积、灰质体积和白质体积的组内相关系数分别为0.69、0.51和0.60。
  • FreeSurfer仅全脑体积组内相关系数为0.63。
  • SPM的基于体素形态学分析在薄层图像与插值厚层图像上,大多数脑区组内相关系数为0.37-0.98,白质88.47%(306924/346916体素)和灰质80.35%(377282/469502体素)的脑区达到尚可至优异。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Purpose To validate the use of thick-section clinically acquired magnetic resonance (MR) imaging data for estimating total brain volume (TBV), gray matter (GM) volume (GMV), and white matter (WM) volume (WMV) by using three widely used automated toolboxes: SPM ( www.fil.ion.ucl.ac.uk/spm/ ), FreeSurfer ( surfer.nmr.mgh.harvard.edu ), and FSL (FMRIB software library; Oxford Centre for Functional MR Imaging of the Brain, Oxford, England, https://fsl.fmrib.ox.ac.uk/fsl ). Materials and Methods MR images from a clinical archive were used and data were deidentified. The three methods were applied to estimate brain volumes from thin-section research-quality brain MR images and routine thick-section clinical MR images acquired from the same 38 patients (age range, 1-71 years; mean age, 22 years; 11 women). By using these automated methods, TBV, GMV, and WMV were estimated. Thin- versus thick-section volume comparisons were made for each method by using intraclass correlation coefficients (ICCs). Results SPM exhibited excellent ICCs (0.97, 0.85, and 0.83 for TBV, GMV, and WMV, respectively). FSL exhibited ICCs of 0.69, 0.51, and 0.60 for TBV, GMV, and WMV, respectively, but they were lower than with SPM. FreeSurfer exhibited excellent ICC of 0.63 only for TBV. Application of SPM's voxel-based morphometry on the modulated images of thin-section images and interpolated thick-section images showed fair to excellent ICCs (0.37-0.98) for the majority of brain regions (88.47% [306924 of 346916 voxels] of WM and 80.35% [377 282 of 469 502 voxels] of GM). Conclusion Thick-section clinical-quality MR images can be reliably used for computing quantitative brain metrics such as TBV, GMV, and WMV by using SPM. © RSNA, 2017 Online supplemental material is available for this article.

来源:Radiology,查看原文。版权归原作者所有。

引用

  1. Adduru VR, Michael AM, Helguera M, et al. Leveraging Clinical Imaging Archives for Radiomics: Reliability of Automated Methods for Brain Volume Measurement. Radiology. 2017. PMID: 28448234. doi:10.1148/radiol.2017161928

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。