腕关节超高分辨率C臂CT关节造影:辐射剂量与图像质量对比传统多排CT
本模体与尸体研究比较了超高分辨率1×1合并C臂CT(CACT)与标准16排CT(MDCT)腕关节造影的有效辐射剂量和图像质量。MDCT有效剂量与标准CACT相当(4.3μSv对3.7μSv);高对比分辨率CACT2最佳,CACT1次之,MDCT最差;低对比分辨率从CACT2到MDCT增高(P<0.001);图像质量评分CACT2最佳(1.3±0.5),CACT1次之(1.9±0.6),MDCT最差(3.5±0.6);仅CACT报告非妥协性伪影。作者认为超高分辨率C臂CT腕关节造影在图像质量和工作流程方面有潜力优于MDCT,代价是图像伪影轻度增加,而两者辐射剂量均较低。
要点
- 超高分辨率C臂CT(CACT2)高对比分辨率和图像质量评分均优于常规16排MDCT。
- CACT2低对比分辨率低于MDCT(P<0.001),但仅CACT出现非妥协性伪影。
- MDCT与CACT有效剂量相当(4.3μSv对3.7μSv),CACT未增加辐射剂量。
核心结论
- 模体与尸体研究提示,超高分辨率C臂CT腕关节造影在图像质量和工作流程方面有潜力优于MDCT,代价是图像伪影轻度增加,而两者辐射剂量均较低。
研究设计与样本
- 模体与尸体研究;10具人尸体腕关节,三间隙注射稀释碘对比剂及导线模体;比较MDCT与CACT(软核CACT1和锐核CACT2重建);三名放射科医师采用5点Likert量表评估。
关键影像表现/征象
- 高对比分辨率CACT2最佳,CACT1次之,MDCT最差。
- 低对比分辨率从CACT2到MDCT增高(P<0.001)。
- 图像质量评分CACT2为1.3±0.5,CACT1为1.9±0.6,MDCT为3.5±0.6。
- 仅CACT报告非妥协性伪影。
诊断标准与关键数值
- MDCT有效剂量4.3μSv,标准CACT有效剂量3.7μSv。
- 图像质量采用5点Likert量表评估,评分越低图像质量越好。
临床应用与阅片要点
- 超高分辨率C臂CT(锐核重建)可提供优于MDCT的腕关节图像质量和高对比分辨率。
- C臂CT低对比分辨率低于MDCT,阅片时需注意可能出现的非妥协性伪影。
- C臂CT与MDCT辐射剂量均较低且相当(4.3μSv对3.7μSv)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Objective of this phantom and cadaveric study was to compare the effective radiation dose (ED) and image quality (IQ) between C-arm computed tomography (CACT) using an ultra-high resolution 1×1 binning with a standard 16-slice CT (MDCT) arthrography of the wrist.
METHODS: ED was determined with thermoluminescence dosimetry using an anthropomorphic phantom and different patient positions. Imaging was conducted in 10 human cadaveric wrists after tri-compartmental injection of diluted iodinated contrast material and a wire phantom. IQ of MDCT was compared with CACT reconstructed with a soft (CACT1) and sharp (CACT2) kernel. High and low contrast resolution was determined. Three radiologists assessed IQ of wrist structures and occurrence of image artifacts using a 5-point Likert scale.
RESULTS: ED of MDCT was comparable to standard CACT (4.3μSv/3.7μSv). High contrast resolution was best for CACT2, decreased to CACT1 and MDCT. Low contrast resolution increased between CACT2 and MDCT (P<0.001). IQ was best for CACT2 (1.3±0.5), decreased to CACT1 (1.9±0.6) and MDCT (3.5±0.6). Non-compromising artifacts were only reported for CACT.
CONCLUSIONS: The results of this phantom and cadaveric study indicate that ultra-high resolution C-Arm CT arthrography of the wrist bears the potential to outperform MDCT arthrography in terms of image quality and workflow at the cost of mildly increasing image artifacts while radiation dose to the patient is comparably low for both, MDCT and C-Arm CT.
引用
- Werncke T, Sonnow L, Meyer BC, et al. Ultra-high resolution C-Arm CT arthrography of the wrist: Radiation dose and image quality compared to conventional multidetector computed tomography. Eur J Radiol. 2017. PMID: 28267538. doi:10.1016/j.ejrad.2017.02.009
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。