使用单超声线阵换能器的多焦点波束形成用于热应变成像

Multi-Focus Beamforming for Thermal Strain Imaging Using a Single Ultrasound Linear Array Transducer.
PubMed 文献 · Ultrasound Med Biol · 2017-03-18

本研究设计了一种可在临床线阵成像换能器上实现的多焦点加热波束,用于超声诱导热应变成像(TSI)。体外实验显示有效加热面积约0.85 cm²、2秒加热升温0.3°C,与有限元模拟吻合;该波束可检测直径1 cm的脂质模拟橡胶包涵体,应变对比度2.3(橡胶内+0.14%应变,明胶背景-0.06%应变),并在1 cm直径人颈动脉内膜剥脱标本中识别脂质组织,与组织学结果一致。

要点

  • 多焦点加热波束可在临床线阵换能器上实现,有效加热面积约0.85 cm²、2秒升温0.3°C。
  • TSI可检出直径1 cm脂质模拟包涵体,应变对比度2.3(包涵体+0.14%,背景-0.06%)。
  • 在1 cm人颈动脉内膜剥脱标本中识别脂质组织,与组织学结果一致。

核心结论

  • 新设计的加热波束可在线阵成像换能器上实现,提供优于既往实现的加热面积和效率,并成功用于体外和离体标本的脂质组织TSI检测。

研究设计与样本

  • 研究论文,评估一种可在临床线阵成像换能器上实现的多焦点加热波束设计,使用组织模拟体模进行体外实验,并与计算机有限元模拟比较,另用1 cm直径人颈动脉内膜剥脱标本验证。

关键影像表现/征象

  • 体外组织模拟体模(无血流)中,新设计有效加热面积约0.85 cm²,2秒加热升温0.3°C,与有限元模拟吻合良好。
  • TSI可从水基明胶背景中检出直径1 cm的脂质模拟橡胶包涵体,应变对比度2.3(橡胶包涵体+0.14%应变,明胶背景-0.06%应变)。
  • 在1 cm直径人颈动脉内膜剥脱标本中识别出脂质组织,与组织学结果一致。

诊断标准与关键数值

  • 脂质模拟橡胶包涵体直径1 cm时,TSI应变对比度为2.3(包涵体+0.14%应变,明胶背景-0.06%应变)。
  • 有效加热面积约0.85 cm²,2秒加热升温0.3°C。

临床应用与阅片要点

  • 本研究为体外和离体实验,原文未据此给出临床阅片或报告推荐。

局限性

  • 原文指出TSI在临床实现面临运动伪影和位移追踪准确性挑战。
  • 原文指出TSI加热能力有限,导致热应变信噪比低。
  • 原文指出TSI视野有限。
  • 原文指出当前TSI系统使用分离的加热和成像换能器,需物理对准加热与成像波束,装置笨重,限制在体操作。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Ultrasound-induced thermal strain imaging (TSI) has been used successfully to identify lipid- and water-based tissues in atherosclerotic plaques in some research settings. However, TSI faces several challenges to be realized in clinics. These challenges include motion artifacts and displacement tracking accuracy, as well as limited heating capability, which contributes to low thermal strain signal-to-noise ratio, and a limited field of view. Our goal was to address the challenge in heating tissue in TSI. Current TSI systems use separate heating and imaging transducers, which require physical alignment of the heating and imaging beams and result in a bulky setup that limits in vivo operation. We evaluated a new design for heating beams that can be implemented on a linear array imaging transducer and can provide improved heating area and efficiency as compared with previous implementations. The heating beams designed were implemented with a clinical linear array imaging transducer connected to a research ultrasound platform. In vitro experiments using tissue-mimicking phantoms with no blood flow revealed that the new design resulted in an effective heating area of approximately 0.85 cm2 and a 0.3°C temperature rise in 2 s of heating, which compared well with in silico finite-element simulations. With the new heating beams, TSI was found to be able to detect a lipid-mimicking rubber inclusion with a diameter of 1 cm from the water-based gelatin background, with a strain contrast of 2.3 (+0.14% strain in the rubber inclusion and -0.06% strain in the gelatin background). Lastly, lipid-based tissue in a 1-cm-diameter human carotid endarterectomy (CEA) sample was identified in good agreement with histology.