文献速递骨肌MR

定性3-T质子磁共振波谱用于肌肉骨骼肿瘤表征:诊断性能与适应证的更新

Qualitative 3-T Proton MR Spectroscopy for the Characterization of Musculoskeletal Neoplasms: Update on Diagnostic Performance and Indications.
PubMed 文献 · AJR Am J Roentgenol · 2017-03-16

该前瞻性研究纳入74例患者76个经组织学证实的肌肉骨骼肿瘤,在3-T MRI上行质子磁共振波谱,视觉评估3.2 ppm处胆碱峰有无,以胆碱峰存在提示恶性。两种软件诊断性能相似(κ=0.97);定性1H-MRS无法鉴别良恶性骨肿瘤(最佳敏感度50.0%、特异度61.5%),对软组织病变敏感度较好但特异度仍低(最佳敏感度82.4%、特异度64.3%),应用谱质标准时排除35.4%的软组织谱。结论:定性3-T 1H-MRS不能用于骨肿瘤表征,对软组织病变敏感度高但特异度低。

要点

  • 定性3-T 1H-MRS以3.2 ppm胆碱峰存在提示恶性,不能用于骨肿瘤良恶性鉴别。
  • 对软组织病变最佳敏感度82.4%、特异度64.3%,特异度仍低。
  • 应用谱质标准评估软组织肿瘤时排除35.4%的谱。

核心结论

  • 定性3-T 1H-MRS不能用于骨肿瘤的表征。
  • 该技术对软组织病变表征有良好敏感度但特异度低。

研究设计与样本

  • 前瞻性纳入74例患者76个经组织学证实的肌肉骨骼肿瘤,使用3-T MRI扫描仪行质子磁共振波谱。
  • 谱分别用设备厂商提供的常规MRI软件和专用独立MRS软件分析,视觉评估3.2 ppm处胆碱峰有无,以胆碱峰存在提示恶性。
  • 考虑肿瘤来源和谱质量对诊断性能的影响。

关键影像表现/征象

  • 视觉评估3.2 ppm处胆碱峰的有无,以胆碱峰存在提示恶性。

诊断标准与关键数值

  • 骨肿瘤:无论是否应用质量 criteria,定性1H-MRS均不能鉴别良恶性,最佳敏感度50.0%、特异度61.5%。
  • 软组织病变:最佳敏感度82.4%、特异度64.3%。
  • 两种软件诊断性能相似(κ=0.97)。
  • 应用谱质标准评估软组织肿瘤时排除35.4%的受评估谱。

临床应用与阅片要点

  • 定性3-T 1H-MRS不能用于骨肿瘤表征。
  • 对软组织病变该技术有良好敏感度但特异度低。

局限性

  • 定性3-T 1H-MRS对软组织病变特异度仍低(最佳特异度64.3%)。
  • 应用谱质标准评估软组织肿瘤时排除35.4%的受评估谱。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: The objective of our study was to evaluate the diagnostic performance of qualitative 3-T proton MR spectroscopy (1H-MRS) for the characterization of musculoskeletal neoplasms.

SUBJECTS AND METHODS: Proton MRS studies of 74 patients (76 lesions) with a histologically confirmed musculoskeletal neoplasm or neoplasms were prospectively included in this study. All studies were performed using a 3-T MRI scanner. Spectra were analyzed with conventional MRI software provided by the fabricant and with dedicated independent MRS software. Spectra were evaluated visually for the presence or absence of a choline peak at 3.2 ppm. The presence of a choline peak was considered indicative of malignancy. The influences of tumor origin and spectral quality on diagnostic performance were considered.

RESULTS: Diagnostic performance was similar with both software used (κ = 0.97). Qualitative 1H-MRS failed to differentiate benign from malignant bone tumors regardless of the application of quality criteria (best sensitivity and specificity, 50.0% and 61.5%, respectively). Diagnostic performance was better with soft-tissue lesions, but the specificity remained low (best sensitivity and specificity, 82.4% and 64.3%, respectively). The application of spectral quality criteria for the evaluation of soft-tissue tumors led to the exclusion of 35.4% of the evaluated spectra.

CONCLUSION: Qualitative 3-T 1H-MRS cannot be used to characterize bone tumors. This technique has good sensitivity but poor specificity for the characterization of soft-tissue lesions.

来源:AJR Am J Roentgenol,查看原文。版权归原作者所有。

引用

  1. Gondim Teixeira PA, Ledrich M, Kauffmann F, et al. Qualitative 3-T Proton MR Spectroscopy for the Characterization of Musculoskeletal Neoplasms: Update on Diagnostic Performance and Indications. AJR Am J Roentgenol. 2017. PMID: 28301222. doi:10.2214/AJR.16.17285

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。