文献速递乳腺MR

仰卧位乳腺磁共振成像采用呼吸触发技术

Supine Breast MRI Using Respiratory Triggering.
PubMed 文献 · Acad Radiol · 2017-02-27

本研究在3T下比较仰卧位呼吸触发MRI与仰卧位非触发MRI及常规俯卧位非触发MRI,每组10名志愿者(含1例患者)。仰卧位触发MRI图像质量指标(SNR、CNR、锐度)较仰卧位非触发MRI改善,SNR和CNR较俯卧位MRI改善,但采集时间从56.5秒延长至平均306秒。

要点

  • 仰卧位呼吸触发乳腺MRI可减少呼吸运动伪影,提高图像质量。
  • 与仰卧位非触发MRI相比,触发后SNR、CNR和图像锐度均改善(P<0.01)。
  • 与俯卧位非触发MRI相比,仰卧位触发MRI的SNR和CNR改善,但图像锐度无显著差异(P=0.88)。

核心结论

  • 应用呼吸触发技术可在类似后续治疗体位下获得高质量仰卧位乳腺MRI,图像质量较仰卧位非触发和俯卧位MRI改善,但采集时间增加。

研究设计与样本

  • 前瞻性对比研究,共30名志愿者(含1例患者),每组10名,在3T下分别行仰卧位呼吸触发MRI、仰卧位非触发MRI和俯卧位非触发MRI(16通道乳腺线圈);仰卧位使用32通道胸部线圈置于覆盖物上方以防止乳房变形。

诊断标准与关键数值

  • 仰卧位触发MRI较仰卧位非触发MRI:SNR中位数从11.5(四分位距6.0)升至38.1(29.1),CNR从7.3(3.1)升至32.8(29.7),图像锐度从0.23(0.2)cm降至0.12(0)cm,均P<0.01;仰卧位触发MRI较俯卧位非触发MRI:SNR从14.7(6.8)升至36.2(12.2),CNR从12.6(5.6)升至32.7(12.1),均P<0.01,图像锐度0.10(0)cm vs 0.11(0)cm,P=0.88。

临床应用与阅片要点

  • 仰卧位呼吸触发乳腺MRI可用于需要在治疗体位下成像的患者,但需注意采集时间较非触发序列明显延长(平均306秒,范围120-540秒)。

局限性

  • 本研究每组仅10名志愿者(含1例患者),样本量小。
  • 仰卧位触发MRI采集时间较非触发MRI延长(从56.5秒增至平均306秒)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: This study aims to evaluate if navigator-echo respiratory-triggered magnetic resonance acquisition can acquire supine high-quality breast magnetic resonance imaging (MRI).

MATERIALS AND METHODS: Supine respiratory-triggered magnetic resonance imaging (Trig-MRI) was compared to supine non-Trig-MRI to evaluate breathing-induced motion artifacts (group 1), and to conventional prone non-Trig-MRI (group 2, 16-channel breast coil), all at 3T. A 32-channel thorax coil was placed on top of a cover to prevent breast deformation. Ten volunteers were scanned in each group, including one patient. The acquisition time was recorded. Image quality was compared by visual examination and by calculation of signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and image sharpness (IS).

RESULTS: Scan time increased from 56.5 seconds (non-Trig-MRI) to an average of 306 seconds with supine Trig-MRI (range: 120-540 seconds). In group 1, the median values (interquartile range) of SNR, CNR, and IS improved from 11.5 (6.0), 7.3 (3.1), and 0.23 (0.2) cm on supine non-Trig-MRI to 38.1 (29.1), 32.8 (29.7), and 0.12 (0) cm (all P < 0.01) on supine Trig-MRI. All qualitative image parameters in group 1 improved on supine Trig-MRI (all P < 0.01). In group 2, SNR and CNR improved from 14.7 (6.8) and 12.6 (5.6) on prone non-Trig-MRI to 36.2 (12.2) and 32.7 (12.1) (both P < 0.01) on supine Trig-MRI. IS was similar: 0.10 (0) cm vs 0.11 (0) cm (P = 0.88).

CONCLUSIONS: Acquisition of high-quality supine breast MRI is possible when respiratory triggering is applied, in a similar setup as during subsequent treatment. Image quality improved when compared to supine non-triggered breast MRI and prone breast MRI, but at the cost of increased acquisition time.

来源:Acad Radiol,查看原文。版权归原作者所有。

引用

  1. Janssen NNY, Ter Beek LC, Loo CE, et al. Supine Breast MRI Using Respiratory Triggering. Acad Radiol. 2017. PMID: 28256441. doi:10.1016/j.acra.2017.01.003

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。