TIPS植入时镍钛支架扩张不足:解决方案还是错觉?

The underdilation of nitinol stents at TIPS implantation: Solution or illusion?
PubMed 文献 · Eur J Radiol · 2017-02-01

植入时所有支架均明显扩张不足,仅达标称直径的76-92%,最小直径均为8mm,与标称直径及球囊尺寸无关。随访中支架向标称直径扩张,提示扩张不足的实用性存疑。

要点

  • TIPS植入时支架最小直径均为8mm,与标称直径(8、9、10mm)无关
  • 支架最小直径与所用球囊尺寸(8、9、10mm)无关
  • 随访中支架向标称直径扩张,扩张不足的实用性存疑

核心结论

  • TIPS植入时周围组织顺应性主要决定支架直径,标称尺寸或扩张球囊影响很小,因此无法准确调整目标压力梯度;随访中支架向标称直径扩张,对扩张不足的实用性提出质疑。

关键影像表现/征象

  • 植入时所有支架均明显扩张不足,仅达标称直径的76-92%
  • 最小测量直径均为8mm,与标称直径(8、9、10mm)无关
  • 10mm支架的最小直径与所用8、9或10mm球囊无关
  • 平均随访12.7±17.8个月(中位3个月,范围1-81)内支架扩张0.5-1.6mm,取决于标称支架尺寸(8、9、10mm)和原发扩张不足程度

诊断标准与关键数值

  • 最大测量直径与标称直径的Pearson相关系数r=0.952(p<0.001),显示测量准确
  • Viatorr支架与裸支架之间无显著差异

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: This study investigates the behaviour of self-expanding nitinol stents at the time of TIPS-implantation and thereafter.

METHODS: Hundred consecutive patients with cirrhosis receiving a TIPS revision were included. The smallest stent diameter was measured radiologically immediately after implantation and before shunt revision. Accuracy of the measurement was assessed by comparing the nominal stent diameter with the largest stent diameter measured at the time of revision.

RESULTS: Pearson correlation between largest measured and nominal diameters was excellent (r=0.952, p<0.001) showing that measurements are accurate. At TIPS implantation all stents were markedly underdilated reaching only 76-92% of their nominal diameter. Smallest measured diameters were similar (8mm) irrespective of the nominal diameter (8, 9, 10mm) of the stent. In addition, smallest diameters of 10mm stents were similar irrespective whether 8, 9 or 10mm balloons were used. During a mean follow-up of 12.7±17.8months (median 3 months, range 1-81) stents expanded by 0.5-1.6mm dependent on the nominal stent size (8, 9, 10mm) and the grade of primary underdilation. No significant difference was found between Viatorr and bare stents.

CONCLUSIONS: At TIPS-implantation, the compliance of the surrounding tissue predominantly determines the stent diameter. The nominal size of the stent or the dilatation balloon has little influence. Accurate adjustment of a desired pressure gradient is, therefore, not possible. During follow-up, stents expand towards their nominal diameter questioning the usefulness of underdilation.