前瞻性心电触发心脏CT联合低剂量对比剂、低管电压与自适应统计迭代重建用于复杂先天性心脏病婴儿
前瞻性纳入44例复杂先天性心脏病婴儿(年龄8.06±4.33月,体重7.31±1.36 kg),采用80 kVp、120 mA前瞻性心电触发心脏CT,对比剂为碘克沙醇(270 mg碘/ml),70% ASIR重建;有效剂量0.55±0.10 mSv,碘负荷3.95±0.73 g碘,图像质量均可诊断(平均评分4.52±0.38,kappa=0.75),与32例手术结果对照,心外和心内缺损诊断准确率分别为97%和88%。
要点
- 婴儿复杂先心病前瞻性心电触发心脏CT有效剂量0.55±0.10 mSv、碘负荷3.95±0.73 g碘。
- 图像质量均可满足诊断,平均评分4.52±0.38,两名阅片者一致性良好(kappa=0.75)。
- 对照32例手术结果,心外缺损诊断准确率97%,心内缺损诊断准确率88%。
核心结论
- 前瞻性心电触发心脏CT采用80 kVp、低浓度碘对比剂(270 mg碘/ml)和70% ASIR重建,在降低对比剂剂量和辐射剂量的同时,为婴儿复杂先天性心脏病提供优良图像质量和准确诊断。
研究设计与样本
- 前瞻性研究连续纳入44例复杂先天性心脏病婴儿(男19例、女25例,年龄8.06±4.33月,体重7.31±1.36 kg),采用80 kVp、120 mA前瞻性心电触发心脏CT,对比剂为碘克沙醇(270 mg碘/ml),70% ASIR重建;以手术结果为参考标准评估其中32例的诊断准确率。
关键影像表现/征象
- 所有图像均可满足诊断,两名独立阅片者平均评分4.52±0.38,阅片者间一致性良好(kappa=0.75)。
- 脂肪组织图像噪声16.24±1.42 HU,主动脉对比噪声比21.90±7.10。
诊断标准与关键数值
- 与32例手术结果对照,CT诊断心外缺损准确率97%,诊断心内缺损准确率88%。
- 图像质量采用五分法评分,两名独立阅片者评估并计算阅片者间变异。
临床应用与阅片要点
- 对婴儿复杂先心病可采用80 kVp、120 mA前瞻性心电触发心脏CT,配合低浓度碘对比剂(270 mg碘/ml)与70% ASIR重建。
- CT对心外缺损诊断准确率97%,高于对心内缺损的88%,报告心内结构时需结合其他检查综合判断。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
AIM: To demonstrate the clinical value of prospective electrocardiography (ECG)-triggered cardiac computed tomography (CT) with low concentration contrast medium, low tube voltage, and adaptive statistic iterative reconstruction (ASIR) to reduce both radiation and contrast dose in examining infants with complex congenital heart disease (CHD).
MATERIALS AND METHODS: Forty-four consecutive infants (19 male, 25 female, age: 8.06±4.33 months, weight: 7.31±1.36 kg) with complex CHD underwent prospective ECG-triggered low-dose cardiac CT using 80 kVp and 120 mA. The contrast agent was iodixanol (270 mg iodine/ml, Visipaque, GE Healthcare, Co. Cork, Ireland). Cardiac CT images were reconstructed with 70% ASIR. The quantitative CT image quality was assessed by image noise in adipose tissue and contrast-to-noise ratio (CNR) in the aorta. The qualitative image analysis was performed on a five-point grading scale by two independent reviewers and interobserver variability was calculated. The results of 32 CT examinations were also compared with the available surgical results for diagnostic accuracy evaluation.
RESULTS: The effective dose was 0.55±0.10 mSv for the patient population. The iodine load was 3.95±0.73 g iodine. Image noise in adipose tissue was 16.24±1.42 HU and CNR in aorta was 21.90±7.10. All images were acceptable for diagnosis with an average score of 4.52±0.38 and good agreement between reviewers (kappa=0.75). Compared to the surgery results in 32 cases, CT was 97% and 88% accurate diagnosing extracardiac and intracardiac defects, respectively.
CONCLUSION: Prospective ECG-triggered cardiac CT using 80 kVp, low-concentration iodinated contrast agent (270 mg iodine/ml) and 70% ASIR reconstruction provides excellent image quality and accurate diagnosis for complex congenital heart disease in infants with reduced contrast medium dose and low radiation dose.
引用
- Wang SY, Gao W, Zhong YM, et al. Prospective ECG-triggering cardiac CT for infants with complex congenital heart disease using low-dose contrast medium, low tube voltage, and adaptive statistical iterative reconstruction. Clin Radiol. 2017. PMID: 28267987. doi:10.1016/j.crad.2017.01.017
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。