多排CT中患者体位对CT值及图像噪声均匀性的影响
该研究用人体模型在偏离等中心不同垂直距离下测量上胸部、中胸部及腹部ROI的CT值与噪声,并在20316例临床腹部CT中统计垂直错位程度。结果显示偏离等中心10 cm时部分ROI的CT值绝对变化超过20 HU,10 cm低于等中心时各区域噪声均匀性变化超过两倍;临床扫描中错位超过1、2、4、6 cm者分别占41%、19%、1.9%、0.3%。作者认为患者明显错位时应谨慎解读CT值测量。
要点
- 偏离等中心10 cm时部分ROI的CT值绝对变化超过20 HU,胸部与腹部区域差异重要。
- 10 cm低于等中心时所有区域噪声均匀性变化超过两倍。
- 临床腹部CT中错位超过1 cm者占41%,超过2 cm者占19%。
核心结论
- 患者明显错位时放射科医生应谨慎对待CT值测量;本研究所见衰减值的大幅变化足以值得进一步研究。
研究设计与样本
- 人体模型研究:模型在相对等中心不同垂直方向错位下接受成像,用ROI测量上胸部、中胸部及腹部CT值与噪声;另收集20316例临床腹部CT的垂直错位数据,用箱线图分析确定患者体位范围。
关键影像表现/征象
- 模型偏离等中心10 cm时,部分ROI的CT值绝对变化超过20 HU,胸部与腹部区域间差异重要。
- 模型位于等中心下方10 cm时,所有区域的噪声均匀性变化超过两倍。
临床应用与阅片要点
- 临床腹部CT中垂直错位超过1、2、4、6 cm的患者分别占41%、19%、1.9%、0.3%。
- 患者明显错位时,放射科医生应谨慎对待CT值测量。
局限性
- 原文结论指出本研究所见衰减值的大幅变化足以值得进一步研究。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Many algorithms for clinical decision making rely on assessment of the CT number (expressed as Hounsfield units); however, to our knowledge, few, if any, studies have addressed how CT numbers change as a function of patient positioning within the scanner.
MATERIALS AND METHODS: An anthropomorphic phantom underwent imaging with varying amounts of vertical orientation misalignment with respect to isocenter. CT number and noise were measured using ROIs in the upper thorax, mid thorax, and abdomen. The degree of noise nonuniformity and changes in the CT number were assessed by comparing values obtained in the anterior versus posterior ROIs. To add clinical relevance, data on vertical mispositioning were collected from 20,316 clinical abdominal CT scans. Box-and-whisker plot analysis was used to identify the range of patient positioning.
RESULTS: Absolute CT number changes of more than 20 HU were observed for some ROIs at phantom positions of 10 cm from isocenter, with important differences noted between the thoracic and abdominal regions. Noise uniformity varied by more than twofold for all regions at 10 cm below isocenter. On clinical CT examinations, off-centering of more than 1, 2, 4, and 6 cm occurred for 41%, 19%, 1.9%, and 0.3% of patients, respectively.
CONCLUSION: Radiologists should treat CT number measurements with caution when patients are grossly mispositioned in the scanner. The substantial changes in attenuation values shown in the present study are large enough to warrant further investigation.