省级诊断参考水平与监测实践有助于降低CT检查患者剂量
本研究建立省级诊断参考水平(DRL),并评估其能否降低最常见CT检查的患者辐射剂量。对头部、胸部、低剂量胸部、腹部和盆腔、胸部腹部和盆腔CT各纳入15名平均体重患者(70±20 kg),以患者剂量分布的第75百分位计算DRL,对超过DRL的医院建议降低剂量并再次调查。实施DRL后通过协议变更使患者剂量降低最多41%,图像质量未出现退化。
要点
- 省级DRL以患者剂量分布第75百分位计算,超过DRL的医院建议降低剂量。
- 低剂量胸部CT各扫描仪间平均剂量差异最大,超过五倍。
- 实施DRL后通过协议变更使患者剂量降低最多41%,图像质量无退化。
核心结论
- 建立省级DRL可在不导致图像质量退化的情况下有效降低患者剂量。
研究设计与样本
- 前瞻性多中心调查,初始调查纳入1185名患者,协议优化后追加调查180名患者;每种CT协议各纳入15名平均体重患者(70±20 kg);以患者剂量分布第75百分位计算省级DRL;采用单因素方差分析比较扫描仪间剂量差异,Pearson相关系数评估剂量与扫描仪使用年限及探测器排数的相关性;由经验丰富的放射科医师对腹部和胸部检查样本随机盲法评阅并分级诊断图像质量。
诊断标准与关键数值
- 省级DRL定义为患者剂量分布的第75百分位;各扫描仪剂量分布均值差异在所有检查中均具有统计学意义(p < 0.05);低剂量胸部CT扫描仪间平均剂量值差异超过五倍;剂量与扫描仪使用年限或探测器排数之间相关性极弱;图像质量各评分类别均无统计学显著差异,腹部成像噪声类别除外;实施DRL后通过协议变更使患者剂量降低最多41%。
临床应用与阅片要点
- 对剂量超过省级DRL的医院建议降低剂量,随后再次调查以监测效果。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The purpose of this study is to establish provincial diagnostic reference levels (DRLs) and to determine whether this process may help reduce the patient radiation dose from the most frequently performed CT examinations.
MATERIALS AND METHODS: We investigated the following CT examinations: head, chest, low-dose chest, abdomen and pelvis, and chest, abdomen, and pelvis examinations. The sample for each protocol included 15 patients of average body weight (mean [± SD], 70 ± 20 kg). The differences in dose between scanners were evaluated using one-way ANOVA. Correlations between dose, scanner age, and the number of detector rows were assessed using the Pearson correlation coefficient. A sample of abdominal and chest examinations were randomized and blinded for review by experienced radiologists who graded diagnostic image quality. Provincial DRLs were calculated as the 75th percentile of patient dose distributions. For hospitals with doses exceeding the DRLs, dose reduction was recommended, followed by another survey.
RESULTS: The initial survey included data of 1185 patients, and an additional 180 patients were surveyed after protocol optimization. The differences between the mean values of the dose distributions from each scanner were statistically significant (p < 0.05) for all examinations. The variation was greatest for low-dose chest CT, with a greater than fivefold difference in the mean dose values noted between scanners. A very weak correlation was found between dose and scanner age or the number of detector rows. Analysis of image quality revealed no statistically significant differences in any scoring categories, with the exception of the noise category in abdominal imaging. Implementation of the DRLs allowed a reduction in patient dose of up to 41% as a result of a protocol change.
CONCLUSION: Establishing provincial DRLs allows an effective reduction in patient dose without resulting in degradation of image quality.