基于器官的新型管电流调制方法降低头部CT辐射剂量:体模研究中图像质量与眼晶状体辐射剂量的评估

New organ-based tube current modulation method to reduce the radiation dose during computed tomography of the head: evaluation of image quality and radiation dose to the eyes in the phantom study.
PubMed 文献 · Radiol Med · 2017-03-24

本研究在体模中设计了一种基于器官的新型管电流调制方法(NOB-TCM),在眼前部90°径向弧内将管电流降低30%且不增加其余弧区管电流;与参考扫描及五种减剂量方法比较,NOB-TCM使眼剂量降至参考扫描的25.88%,平均SNR降至8.02%,平均品质因数较参考扫描增加13.39%,作者认为NOB-TCM在无明显图像质量损失下降低眼部辐射暴露,是头部CT的优越方案。

要点

  • NOB-TCM在眼前部90°径向弧内将管电流降低30%,其余径向弧管电流不增加。
  • NOB-TCM使眼剂量降至参考扫描的25.88%,优于铅眼镜的44.53%和铋屏蔽的36.91%。
  • NOB-TCM平均SNR降至参考扫描的8.02%,前部ROI的SNR降至11.89%。

核心结论

  • NOB-TCM是头部CT(包括眶区)的优越方案,可降低辐射暴露且无明显图像质量损失。

研究设计与样本

  • 体模研究,比较参考扫描与五种减剂量方法对剂量、实用性和图像质量的影响。
  • 扫描方法包括纵向TCM、角度TCM和NOB-TCM三种TCM扫描,以及铋片和铅眼镜两种屏蔽扫描。

关键影像表现/征象

  • NOB-TCM在眼前部90°径向弧内将管电流降低30%,其余径向弧管电流不增加。
  • NOB-TCM平均SNR降至参考扫描的8.02%,前部ROI的SNR降至11.89%。
  • 铋屏蔽平均SNR降至参考扫描的32.95%,前部ROI的SNR降至87.89%。
  • 铅眼镜平均SNR降至参考扫描的28.36%。

诊断标准与关键数值

  • NOB-TCM眼剂量降至参考扫描的25.88%,铅眼镜为44.53%,铋屏蔽为36.91%。
  • NOB-TCM平均品质因数较参考扫描增加13.39%,纵向TCM增加11.7%。

临床应用与阅片要点

  • 作者认为NOB-TCM是头部CT(包括眶区)的优越方案,可降低辐射暴露且无明显图像质量损失。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

A new organ-based tube current modulation (NOB-TCM) method was designed with the intent to decrease tube current by 30% over a prescribed 90° radial arc across the anterior aspect of the radiosensitive organ, without increasing tube current in the remaining radial arc. We compared a reference scan and five other dose-reducing methods with regard to effects on dose, practicality, and image quality to determine the most effective method for the reduction of the radiation dose to the eyes during CT examinations of the head. We compared the radiation doses to the eyes and physical image quality in different regions of interest for TCM and shielding scans. Three types of TCM scans were performed: longitudinal TCM, angular TCM, and NOB-TCM. A bismuth sheet and lead goggles were each applied for the shielding scan. Relative to the reference scan, the dose to the eye was reduced to 25.88% with NOB-TCM, 44.53% with lead goggles, and 36.91% with a bismuth shield. Relative to the reference scan, the mean signal-to-noise ratio (SNR) was decreased to 8.02% with NOB-TCM, 28.36% with lead goggles, and 32.95% with the bismuth shield. The SNR of the anterior region of interest was decreased to 11.89% with NOB-TCM and 87.89% with the bismuth shield. The average figure of merit was increased by 11.7% with longitudinal TCM and 13.39% with NOB-TCM, compared with the reference scan. NOB-TCM is a superior solution for head CT, including the orbital area, due to the reduction in radiation exposure without significant loss in image quality.