脑CT诊断小脑后颅窝卒中:专为脑CT设计的迭代重建对检出性能的影响
研究纳入37例后颅窝缺血性卒中患者和37例无卒中对照,比较滤波反投影(FBP)与专为脑CT设计的迭代模型重建(IMR-neuro)的图像质量和低衰减检出性能;IMR-neuro噪声更低、梗死与非梗死区CNR差值更高,ROC曲线下面积显著更高(0.90 vs 0.86,p=0.04)。
要点
- IMR-neuro较FBP显著降低脑桥图像噪声(2.2±0.5 vs 5.1±0.9 HU,p<0.01)。
- IMR-neuro梗死与非梗死区CNR差值显著高于FBP(2.2±1.7 vs 4.0±3.6,p<0.01)。
- IMR-neuro平均ROC曲线下面积显著高于FBP(0.90 vs 0.86,p=0.04)。
核心结论
- IMR-neuro图像质量更好,并改善缺血性卒中患者的低衰减检出。
研究设计与样本
- 研究纳入37例后颅窝缺血性卒中患者和37例无卒中对照,比较FBP与IMR-neuro重建的轴位图像,采用ROC曲线技术分析低衰减检出性能。
关键影像表现/征象
- IMR-neuro脑桥图像噪声显著低于FBP(2.2±0.5 vs 5.1±0.9 HU,p<0.01)。
- IMR-neuro梗死与非梗死区CNR差值显著高于FBP(2.2±1.7 vs 4.0±3.6,p<0.01)。
诊断标准与关键数值
- IMR-neuro平均ROC曲线下面积显著高于FBP(0.90 vs 0.86,p=0.04)。
临床应用与阅片要点
- IMR-neuro可改善后颅窝低对比病灶的检出,有助于后颅窝缺血性卒中的诊断。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: In this study, we aimed to determine whether iterative model reconstruction designed for brain CT (IMR-neuro) would improve the accuracy of posterior fossa stroke diagnosis on brain CT.
METHODS: We enrolled 37 patients with ischaemic stroke in the posterior fossa and 37 patients without stroke (controls). Using axial images reconstructed using filtered back-projection (FBP) and IMR-neuro, we compared the CT numbers in infarcted areas, image noise in the pons, and contrast-to-noise ratios (CNRs) of infarcted and non-infarcted areas on scans subjected to IMR-neuro and FBP. To analyse the performance of hypo-attenuation detection, we used receiver-operating characteristic (ROC) curve techniques.
RESULTS: The image noise was significantly lower (2.2 ± 0.5 vs. 5.1 ± 0.9 Hounsfield units, p < 0.01) and the difference in CNR between the infarcted and non-infarcted areas was significantly higher with IMR-neuro than with FBP (2.2 ± 1.7 vs. 4.0 ± 3.6, p < 0.01). Furthermore, the average area under the ROC curve was significantly higher with IMR-neuro (0.90 vs. 0.86 for FBP, p = 0.04).
CONCLUSION: IMR-neuro yielded better image quality and improved hypo-attenuation detection in patients with ischaemic stroke.
KEY POINTS: • Iterative model reconstruction of brain CT data can facilitate the diagnosis of ischaemic stroke. • IMR improved the detectability of low-contrast lesions in the posterior fossa. • IMR-neuro yielded better image quality and improved observer performance.
引用
- Inoue T, Nakaura T, Yoshida M, et al. Diagnosis of small posterior fossa stroke on brain CT: effect of iterative reconstruction designed for brain CT on detection performance. Eur Radiol. 2017. PMID: 28275901. doi:10.1007/s00330-017-4773-4
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。