基于320排多探测器计算机断层扫描的脑肿瘤分割与灌注参数定量半自动体积软件的验证研究

A semi-automated volumetric software for segmentation and perfusion parameter quantification of brain tumors using 320-row multidetector computed tomography: a validation study.
PubMed 文献 · Neuroradiology · 2017-03-24

本研究开发了半自动体积软件NPerfusion,用于在全脑CT灌注图像上分割脑肿瘤并定量灌注参数,纳入29例经病理证实的脑肿瘤患者,与人工分割比较验证。结果显示NPerfusion成功完成所有肿瘤分割,灌注参数定量与人工分割几乎完全一致,Lin一致性相关系数AF为0.9988、EBV为0.9994、PF为0.9976,Bland-Altman分析显示多数差异在一致性界限内。

要点

  • NPerfusion可半自动分割脑肿瘤并定量全肿瘤体积灌注参数。
  • 与人工分割相比,AF、EBV、PF的Lin一致性相关系数分别为0.9988、0.9994、0.9976。
  • Bland-Altman分析显示多数差异在一致性界限内。

核心结论

  • NPerfusion能成功分割脑肿瘤并计算其灌注参数,经与人工分割比较验证,可用于从全脑CT灌注图像计算脑肿瘤的体积灌注参数。

研究设计与样本

  • 验证研究,纳入29例经病理证实的脑肿瘤患者,于2012年8月至2015年2月术前行全脑CT灌注(320排多探测器CT),由商业软件生成灌注参数后经NPerfusion体积定量并与人工分割比较,采用Lin一致性相关系数和Bland-Altman分析验证。

关键影像表现/征象

  • 所有29个脑肿瘤均成功完成分割和全肿瘤体积灌注参数定量,灌注趋势与既往研究一致。

诊断标准与关键数值

  • NPerfusion与人工分割定量灌注参数的一致性:Lin一致性相关系数AF为0.9988、EBV为0.9994、PF为0.9976。
  • Bland-Altman分析显示NPerfusion与商业软件上人工分割之间多数差异在一致性界限内。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: We developed a semi-automated volumetric software, NPerfusion, to segment brain tumors and quantify perfusion parameters on whole-brain CT perfusion (WBCTP) images. The purpose of this study was to assess the feasibility of the software and to validate its performance compared with manual segmentation.

METHODS: Twenty-nine patients with pathologically proven brain tumors who underwent preoperative WBCTP between August 2012 and February 2015 were included. Three perfusion parameters, arterial flow (AF), equivalent blood volume (EBV), and Patlak flow (PF, which is a measure of permeability of capillaries), of brain tumors were generated by a commercial software and then quantified volumetrically by NPerfusion, which also semi-automatically segmented tumor boundaries. The quantification was validated by comparison with that of manual segmentation in terms of the concordance correlation coefficient and Bland-Altman analysis.

RESULTS: With NPerfusion, we successfully performed segmentation and quantified whole volumetric perfusion parameters of all 29 brain tumors that showed consistent perfusion trends with previous studies. The validation of the perfusion parameter quantification exhibited almost perfect agreement with manual segmentation, with Lin concordance correlation coefficients (ρ c) for AF, EBV, and PF of 0.9988, 0.9994, and 0.9976, respectively. On Bland-Altman analysis, most differences between this software and manual segmentation on the commercial software were within the limit of agreement.

CONCLUSIONS: NPerfusion successfully performs segmentation of brain tumors and calculates perfusion parameters of brain tumors. We validated this semi-automated segmentation software by comparing it with manual segmentation. NPerfusion can be used to calculate volumetric perfusion parameters of brain tumors from WBCTP.

来源:Neuroradiology,查看原文。版权归原作者所有。

引用

  1. Chae SY, Suh S, Ryoo I, et al. A semi-automated volumetric software for segmentation and perfusion parameter quantification of brain tumors using 320-row multidetector computed tomography: a validation study. Neuroradiology. 2017. PMID: 28341992. doi:10.1007/s00234-017-1790-6