儿童脑部CT图像质量与辐射剂量:管电压从120 kVp降至80 kVp的影响
纳入30例同时接受80 kVp和120 kVp非增强脑CT的儿童,定量与定性评估图像质量并比较辐射剂量。80 kVp下灰白质衰减和对比噪声比显著增高、噪声相关参数升高、灰白质分化改善但伪影增多,主观噪声和整体图像质量相似,辐射剂量参数显著降低。
要点
- 儿童非增强脑CT采用80 kVp可改善灰白质对比并降低辐射剂量。
- 80 kVp下图像噪声及后颅窝伪影指数高于120 kVp,但主观整体图像质量相似。
- 阅片时需注意80 kVp图像伪影较多,尤其后颅窝区域。
核心结论
- 在儿童非增强脑CT中,管电压从120 kVp降至80 kVp可改善灰白质对比、图像质量相当并降低辐射剂量。
研究设计与样本
- 30例儿童同时接受80 kVp和120 kVp非增强脑CT,进行定量分析(灰白质平均衰减、衰减差、噪声、信噪比、对比噪声比、后颅窝伪影指数)和定性评分(噪声、灰白质分化、伪影、整体图像质量),辐射剂量以CT剂量指数、剂量长度乘积和有效剂量评估。
诊断标准与关键数值
- 80 kVp下灰质和白质平均衰减及对比噪声比显著增高。
- 80 kVp下噪声、信噪比和后颅窝伪影指数高于120 kVp。
- 定性分析中80 kVp图像灰白质分化改善但伪影多于120 kVp图像。
- 主观图像噪声和整体图像质量评分在两种扫描间相似。
- 80 kVp下辐射剂量参数显著低于120 kVp。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Computed tomography (CT) has generated public concern associated with radiation exposure, especially for children. Lowering the tube voltage is one strategy to reduce radiation dose.
OBJECTIVE: To assess the image quality and radiation dose of non-enhanced brain CT scans acquired at 80 kilo-voltage peak (kVp) compared to those at 120 kVp in children.
MATERIALS AND METHODS: Thirty children who had undergone both 80- and 120-kVp non-enhanced brain CT were enrolled. For quantitative analysis, the mean attenuation of white and gray matter, attenuation difference, noise, signal-to-noise ratio, contrast-to-noise ratio and posterior fossa artifact index were measured. For qualitative analysis, noise, gray-white matter differentiation, artifact and overall image quality were scored. Radiation doses were evaluated by CT dose index, dose-length product and effective dose.
RESULTS: The mean attenuations of gray and white matter and contrast-to-noise ratio were significantly increased at 80 kVp, while parameters related to image noise, i.e. noise, signal-to-noise ratio and posterior fossa artifact index were higher at 80 kVp than at 120 kVp. In qualitative analysis, 80-kVp images showed improved gray-white differentiation but more artifacts compared to 120-kVp images. Subjective image noise and overall image quality scores were similar between the two scans. Radiation dose parameters were significantly lower at 80 kVp than at 120 kVp.
CONCLUSION: In pediatric non-enhanced brain CT scans, a decrease in tube voltage from 120 kVp to 80 kVp resulted in improved gray-white matter contrast, comparable image quality and decreased radiation dose.