介入放射学中的患者剂量:法国最常见操作的多中心研究

Patient dose in interventional radiology: a multicentre study of the most frequent procedures in France.
PubMed 文献 · Eur Radiol · 2017-03-13

法国医学物理师学会为满足欧洲指令2013/59/Euratom对介入放射学参考水平(RL)的要求,开展全国回顾性调查,分析神经放射、血管放射和骨关节介入共15种操作,每操作记录10至30例患者的KAP、透视时间、参考空气比释动能和图像数,以第3四分位数计算RL,结果来自36个科室的4600例操作,证实同一操作患者剂量差异大,所提RL低于多数已发表结果。

要点

  • RL按操作的第3四分位数计算,覆盖KAP、参考空气比释动能、透视时间和图像数四项剂量学指标。
  • 脑动脉造影KAP的RL为90 Gy·cm²、透视时间11分钟;胆道引流KAP为35 Gy·cm²、透视时间16分钟。
  • 下肢动脉造影KAP的RL为75 Gy·cm²、透视时间6分钟;椎体成形术KAP为70 Gy·cm²、透视时间11分钟。

核心结论

  • 该全国调查提出了15种介入操作的参考水平,用于基准比较和方案优化;实践与技术的持续演变要求定期更新RL。

研究设计与样本

  • 法国医学物理师学会实施的全国回顾性调查,覆盖36个科室共4600例操作,涉及神经放射、血管放射和骨关节介入15种操作,每种操作记录10至30例患者的KAP、透视时间、参考空气比释动能和图像数,RL取分布的第3四分位数。

诊断标准与关键数值

  • 脑动脉造影KAP的RL为90 Gy·cm²、透视时间11分钟;胆道引流KAP为35 Gy·cm²、透视时间16分钟;下肢动脉造影KAP为75 Gy·cm²、透视时间6分钟;椎体成形术KAP为70 Gy·cm²、透视时间11分钟;上述四种操作按操作复杂度定义RL;全部15种操作均提出四项剂量学指标的RL,且结果低于多数已发表数据。

临床应用与阅片要点

  • 参考水平可用于基准比较和优化协议;本研究的RL按操作复杂度定义,并提出应随实践与技术演变定期更新。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: A national retrospective survey on patient doses was performed by the French Society of Medical physicists to assess reference levels (RLs) in interventional radiology as required by the European Directive 2013/59/Euratom.

METHODS: Fifteen interventional procedures in neuroradiology, vascular radiology and osteoarticular procedures were analysed. Kerma area product (KAP), fluoroscopy time (FT), reference air kerma and number of images were recorded for 10 to 30 patients per procedure. RLs were calculated as the 3rd quartiles of the distributions.

RESULTS: Results on 4600 procedures from 36 departments confirmed the large variability in patient dose for the same procedure. RLs were proposed for the four dosimetric estimators and the 15 procedures. RLs in terms of KAP and FT were 90 Gm.cm2 and 11 mins for cerebral angiography, 35 Gy.cm2 and 16 mins for biliary drainage, 75 Gy.cm2 and 6 mins for lower limbs arteriography and 70 Gy.cm2 and 11 mins for vertebroplasty. For these four procedures, RLs were defined according to the complexity of the procedure. For all the procedures, the results were lower than most of those already published.

CONCLUSIONS: This study reports RLs in interventional radiology based on a national survey. Continual evolution of practices and technologies requires regular updates of RLs.

KEY POINTS: • Delivered dose in interventional radiology depends on procedure, practice and patient. • National RLs are proposed for 15 interventional procedures. • Reference levels (RLs) are useful to benchmark practices and optimize protocols. • RLs are proposed for kerma area product, air kerma, fluoroscopy time and number of images. • RLs should be adapted to the procedure complexity and updated regularly.

来源:Eur Radiol,查看原文。版权归原作者所有。