脑部磁共振成像配准方法的性能评估
本研究在20例患者的3T脑部MR图像上,以T2序列为固定参考,用1175个解剖标志点评估FLAIR、磁敏感加权血管成像和T1增强序列配准前后的欧几里得误差。结果显示不同算法配准后误差和效果比各异,作者建议采用欧几里得误差、欧几里得效果比和统计效果比作为临床配准算法的性能指标。
要点
- 配准前FLAIR、SWAN、T1增强的误差分别为2.07±0.55 mm、2.63±0.62 mm、3.65±2.00 mm。
- 配准后最佳误差:FLAIR 1.55±0.46 mm,SWAN 1.34±0.23 mm,T1增强1.06±0.16 mm。
- 欧几里得效果比和统计效果比可量化比较不同配准算法的性能。
核心结论
- 作者提出欧几里得误差、欧几里得效果比和统计效果比作为临床配准算法的性能指标,可合理比较不同算法。
研究设计与样本
- 在20例患者临床3T扫描仪上采集T2加权、FLAIR、磁敏感加权血管成像和T1增强4种序列,设定1175个解剖标志点,以T2序列为固定参考,应用多种配准算法,计算配准前后欧几里得误差并与标准标志点配准比较。
诊断标准与关键数值
- 欧几里得效果比为配准后剩余欧几里得误差的分数;统计效果比类似但考虑离散度和噪声。配准前误差:FLAIR 2.07±0.55 mm,磁敏感加权血管成像2.63±0.62 mm,T1增强3.65±2.00 mm;配准后最佳误差:FLAIR 1.55±0.46 mm,磁敏感加权血管成像1.34±0.23 mm,T1增强1.06±0.16 mm;对应欧几里得效果比分别为刚性互信息0.493、仿射互信息0.181、商用GE配准0.096,统计效果比分别为0.573、0.352、0.929。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: Clinical brain MR imaging registration algorithms are often made available by commercial vendors without figures of merit. The purpose of this study was to suggest a rational performance comparison methodology for these products.
MATERIALS AND METHODS: Twenty patients were imaged on clinical 3T scanners by using 4 sequences: T2-weighted, FLAIR, susceptibility-weighted angiography, and T1 postcontrast. Fiducial landmark sites (n = 1175) were specified throughout these image volumes to define identical anatomic locations across sequences. Multiple registration algorithms were applied by using the T2 sequence as a fixed reference. Euclidean error was calculated before and after each registration and compared with a criterion standard landmark registration. The Euclidean effectiveness ratio is the fraction of Euclidean error remaining after registration, and the statistical effectiveness ratio is similar, but accounts for dispersion and noise.
RESULTS: Before registration, error values for FLAIR, susceptibility-weighted angiography, and T1 postcontrast were 2.07 ± 0.55 mm, 2.63 ± 0.62 mm, and 3.65 ± 2.00 mm, respectively. Postregistration, the best error values for FLAIR, susceptibility-weighted angiography, and T1 postcontrast were 1.55 ± 0.46 mm, 1.34 ± 0.23 mm, and 1.06 ± 0.16 mm, with Euclidean effectiveness ratio values of 0.493, 0.181, and 0.096 and statistical effectiveness ratio values of 0.573, 0.352, and 0.929 for rigid mutual information, affine mutual information, and a commercial GE registration, respectively.
CONCLUSIONS: We demonstrate a method for comparing the performance of registration algorithms and suggest the Euclidean error, Euclidean effectiveness ratio, and statistical effectiveness ratio as performance metrics for clinical registration algorithms. These figures of merit allow registration algorithms to be rationally compared.