手持超声系统实时计算机辅助检测腰椎解剖结构的成像性能:可行性研究

Imaging Performance of a Handheld Ultrasound System With Real-Time Computer-Aided Detection of Lumbar Spine Anatomy: A Feasibility Study.
PubMed 文献 · Invest Radiol · 2017-01-01

本研究评估了手持超声系统成像性能及自动化腰椎计算机辅助检测(CAD)算法的准确性,对68名体重指数18.5–48 kg/m²的志愿者进行腰椎成像,并以放射科医生的分割作为参考标准。CAD检测硬膜外隙的敏感度为94.2%、特异度为85.5%,检测脊柱中线的敏感度为93.9%、特异度为91.3%。骨增强成像模式的骨对比度较同类手持超声系统提高5.1–10倍,结果证明CAD辅助床旁实时解读腰椎超声图像可行。

要点

  • CAD检测硬膜外隙敏感度94.2%、特异度85.5%,深度误差约±0.5 cm。
  • CAD检测脊柱中线敏感度93.9%、特异度91.3%,横向位置误差约±0.3 cm。
  • 骨增强成像模式骨对比度较同类手持超声系统提高5.1–10倍。

核心结论

  • 研究结果表明,计算机辅助检测用于辅助床旁实时解读腰椎超声图像具有可行性。

研究设计与样本

  • 可行性研究,纳入68名体重指数18.5–48 kg/m²的志愿者,使用作者设计的手持超声系统(具备增强骨成像质量和全自动腰椎CAD),以放射科医生提供的神经轴解剖分割作为参考标准,研究经弗吉尼亚大学机构审查委员会批准。

诊断标准与关键数值

  • CAD检测硬膜外隙的敏感度为94.2%(95%CI 85.1%–98.1%),特异度为85.5%(95%CI 81.7%–88.6%),深度测量误差约±0.5 cm(与从二维超声图像手动测量相比)。
  • CAD检测脊柱中线的敏感度为93.9%(95%CI 85.8%–97.7%),特异度为91.3%(95%CI 83.6%–96.9%),图像内横向位置测量误差约±0.3 cm。
  • 骨增强成像模式产生的骨对比度较同类手持超声成像系统提高5.1–10倍。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The aim of this study was to evaluate the imaging performance of a handheld ultrasound system and the accuracy of an automated lumbar spine computer-aided detection (CAD) algorithm in the spines of human subjects.

MATERIALS AND METHODS: This study was approved by the institutional review board of the University of Virginia. The authors designed a handheld ultrasound system with enhanced bone image quality and fully automated CAD of lumbar spine anatomy. The imaging performance was evaluated by imaging the lumbar spines of 68 volunteers with body mass index between 18.5 and 48 kg/m. The accuracy, sensitivity, and specificity of the lumbar spine CAD algorithm were assessed by comparing the algorithm's results to ground-truth segmentations of neuraxial anatomy provided by radiologists.

RESULTS: The lumbar spine CAD algorithm detected the epidural space with a sensitivity of 94.2% (95% confidence interval [CI], 85.1%-98.1%) and a specificity of 85.5% (95% CI, 81.7%-88.6%) and measured its depth with an error of approximately ±0.5 cm compared with measurements obtained manually from the 2-dimensional ultrasound images. The spine midline was detected with a sensitivity of 93.9% (95% CI, 85.8%-97.7%) and specificity of 91.3% (95% CI, 83.6%-96.9%), and its lateral position within the ultrasound image was measured with an error of approximately ±0.3 cm. The bone enhancement imaging mode produced images with 5.1- to 10-fold enhanced bone contrast when compared with a comparable handheld ultrasound imaging system.

CONCLUSIONS: The results of this study demonstrate the feasibility of CAD for assisting with real-time interpretation of ultrasound images of the lumbar spine at the bedside.

来源:Invest Radiol,查看原文。版权归原作者所有。