数字乳腺断层合成图像中显示器亮度水平如何影响乳腺微钙化和毛刺状病灶的可检测性
研究评估医用显示器校准亮度对数字乳腺断层合成图像中微钙化和毛刺状病灶可检测性的影响,使用4种最大亮度500至1000 cd/m²、最小亮度0.5至1.0 cd/m²的显示器,由1名高年资和2名低年资放射科医生盲法评估48例研究(16例微钙化、16例毛刺状病灶、16例无病灶),结果显示1000 cd/m²显示器较最低亮度显示器病灶可检测性显著更高(P<0.001)。
要点
- 1000 cd/m²亮度显示器较最低亮度显示器对乳腺病灶可检测性显著更高(P<0.001)。
- 两名1000 cd/m²亮度显示器的阅片者间及阅片者与真值间一致性均>0.80,高于较低亮度显示器。
- 阅片者自身可重复性范围为0.75至1.00。
核心结论
- 显示器亮度水平影响数字乳腺断层合成图像中乳腺病灶检测的准确性。
研究设计与样本
- 研究评估4种医用显示器模型,校准最大亮度500至1000 cd/m²、最小亮度0.5至1.0 cd/m²;纳入48例研究(16例微钙化、16例毛刺状病灶、16例无病灶),由1名高年资和2名低年资放射科医生盲法评估。
诊断标准与关键数值
- 阅片者自身可重复性范围为0.75至1.00。
- 阅片者间及阅片者与真值间一致性值>0.80,且两名1000 cd/m²亮度显示器高于较低亮度显示器。
- 1000 cd/m²亮度显示器较最低亮度显示器病灶可检测性显著更高(P<0.001)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: This study evaluates the influence of the calibrated luminance level of medical displays in the detectability of microcalcifications and spiculated lesions in digital breast tomosynthesis images.
MATERIALS AND METHODS: Four models of medical displays with calibrated maximum and minimum luminance, respectively, ranging from 500 to 1000 cd/m2 and from 0.5 to 1.0 cd/m2, were investigated. Forty-eight studies were selected by a senior radiologist: 16 with microcalcifications, 16 with spiculated lesions, and 16 without lesions. All images were anonymized and blindly evaluated by one senior and two junior radiologists. For each study, lesion presence or absence and localization statements, interpretative difficulty level, and overall quality were reported. Cohen's kappa statistic was computed between monitors and within or between radiologists to estimate the reproducibility in correctly identifying lesions; for multireader-multicase analysis, the weighted jackknife alternative free-response receiver operating characteristic statistical tool was applied.
RESULTS: Intraradiologist reproducibility ranged from 0.75 to 1.00. Interreader as well as reader-truth agreement values were >0.80 and higher with the two 1000 cd/m2 luminance displays than with the lower luminance displays for each radiologist. Performances in the detectability of breast lesions were significantly greater with the 1000 cd/m2 luminance displays when compared to the display with the lowest luminance value (P value <0.001).
CONCLUSIONS: Our findings highlight the role of display luminance level on the accuracy of detecting breast lesions.